OZARK Cervical Plate System AA01-43F57VRecalls & reported problems

VB Spine LLC · Plate, 3 Level Size 57 mm · Class 2 · Implant

All models of OZARK Cervical Plate System (54)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10888857402003), received between Apr 12, 2022 and Nov 30, 2022 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component3
  • Device Dislodged or Dislocated1
  • Fracture1
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 30, 2022Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    A PATIENT UNDERWENT REVISION SURGERY TO ADDRESS A FRACTURED LOCKING MECHANISM ON A THREE-LEVEL OZARK PLATE. TWO OZARK SCREWS AT THE LEVEL OF THE FRACTURED LOCKING MECHANISM MIGRATED AND "BEGAN ERODING UNDER THE ESOPHAGUS." THIS REPORT CAPTURES THE THREE-LEVEL OZARK PLATE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 15885673 ↗
  • MalfunctionAug 23, 2022Detachment of Device or Device Component; Device Dislodged or Dislocated; Detachment of Device or Device Component; Device Dislodged or Dislocated

    IT WAS REPORTED THAT AT THE SIX MONTH FOLLOW UP, ONE (1) OZARK SELF-TAPPING, VARIABLE SCREW AT C7 HAS BACKED OUT BEYOND THE LOCKING MECHANISM OF THE OZARK PLATE. THE PATIENT IS DOING WELL, AND A REVISION SURGERY HAS NOT BEEN SCHEDULED. NO ADVERSE CONSEQUENCES, SURGICAL DELAY OR MEDICAL INTERVENTION WERE REPORTED. THIS REPORT CAPTURES THE OZARK PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15279899 ↗
  • MalfunctionApr 12, 2022Material Separation; Detachment of Device or Device Component; Material Separation; Detachment of Device or Device Component

    IT WAS REPORTED THAT APPROXIMATELY 2 MONTHS POST-OPERATIVELY, THE SECURING MECHANISM OF AN OZARK PLATE FAILED CAUSING THE TWO OZARK SCREWS AT C7 TO MIGRATE. REVISION SURGERY HAS NOT BEEN PERFORMED OR SCHEDULED. THIS REPORT CAPTURES PLATE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14084597 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page