Skip to content

Ventlab BT4024Recalls & reported problems

Ventlab, LLC · PEDIATRIC BAG, MASK W/ 24 IN. CORRUGATED TUB,PEEP, P-OFF W/MAN · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Oct 11, 2023Open, Classified

    A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

    Z-0362-2024 · Root cause: Device Design · Recall details · FDA record ↗

Other Mask, Gas, Anesthetic devices

All Mask, Gas, Anesthetic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page