Ventlab HS4011Recalls & reported problems
Ventlab, LLC · 1 LITRE HYPERINFLATION BAG SYSTEM W/ MANOMETER · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10889483110850), received between Jun 1, 2022 and Jun 18, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break1
- Decrease in Pressure1
- Leak/Splash1
- Manufacturing, Packaging or Shipping Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 18, 2025Break; Break
IT WAS REPORTED, THE ENDOTRACHEAL TUBE WAS ATTACHED TO A BAGGED PATIENT WHEN THEY NOTICED PART OF THE CLEAR PLASTIC PART OF THE MANOMETER PIECE WAS BROKEN. IT WAS REPLACED WITH ANOTHER DEVICE; THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22286410 ↗MalfunctionNov 17, 2023Leak/Splash; Leak/Splash
THE BAG WAS LEAKING AIR. THE PATIENT IN THE PACU REQUIRED SUPPORT WITH BREATHING. THE CRNA REQUESTED A BAG VALVE MASK AND RECEIVED A 1L HYPERINFLATION SYSTEM + MANOMETER FROM VENTLAB (SUNMED). AFTER HOOKING UP THE OXYGEN THE CRNA ATTEMPTED VENTILATION AND NOTICED THE BAG LEAKING. ON INSPECTION, THE INFLATION BAG HAD SEVERAL SMALL CUTS THAT PREVENTED GOOD VENTILATION. A REPLACEMENT (SAME PRODUCT) WAS QUICKLY OBTAINED AND THE PATIENT WAS TREATED WITHOUT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18165397 ↗MalfunctionOct 20, 2023
THE PATIENT IN THE PACU REQUIRED SUPPORT WITH BREATHING. THE CRNA REQUESTED A BAG VALVE MASK AND RECEIVED A 1L HYPERINFLATION SYSTEM + MANOMETER FROM VENTLAB (SUNMED). AFTER HOOKING UP THE OXYGEN THE CRNA ATTEMPTED VENTILATION AND NOTICED THE BAG LEAKING. ON INSPECTION, THE INFLATION BAG HAD SEVERAL SMALL CUTS THAT PREVENTED GOOD VENTILATION. A REPLACEMENT (SAME PRODUCT) WAS QUICKLY OBTAINED AND THE PATIENT WAS TREATED WITHOUT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17975566 ↗MalfunctionJun 1, 2022Decrease in Pressure
THE COMMITTEE ON TECHNOLOGY FOR THE ANESTHESIA PATIENT SAFETY FOUNDATION RECEIVED A REPORT REGARDING OUR HYPERINFLATION SYSTEM. DESCRIBING THE DESIGN AS COUNTER-INTUITIVE AND THAT IT DOES NOT CONFORM WITH ISO STANDARD RECOMMENDATIONS. A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE. HOWEVER; THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. THE PROBLEM WAS REMEDIED BY QUICK OBSERVATION THAT TO CLOSE THE VALVE, THE VALVE MUST BE TURNED ANTI-CLOCKWISE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14570258 ↗MalfunctionJun 1, 2022Manufacturing, Packaging or Shipping Problem
THE COMMITTEE ON TECHNOLOGY FOR THE ANESTHESIA PATIENT SAFETY FOUNDATION RECEIVED A REPORT REGARDING OUR HYPERINFLATION SYSTEM DESCRIBING THE DESIGN AS COUNTER-INTUITIVE AND THAT IT DOES NOT CONFORM WITH ISO STANDARD RECOMMENDATIONS. A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE, HOWEVER THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. THE PROBLEM WAS REMEDIED BY QUICK OBSERVATION THAT TO CLOSE THE VALVE, THE VALVE MUST BE TURNED ANTI-CLOCKWISE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14569757 ↗
Other Resuscitator,Manual,Non Self-Inflating devices
- Ventlab HS4000Ventlab, LLC · Resuscitator,Manual,Non Self-Inflating2 reports
- Hyperinflation System 10-55833Mercury Enterprises, Inc. · Resuscitator,Manual,Non Self-Inflating1 report
- Ventlab HS4051Ventlab, LLC · Resuscitator,Manual,Non Self-Inflating1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page