Ventlab HS4011Recalls & reported problems

Ventlab, LLC · 1 LITRE HYPERINFLATION BAG SYSTEM W/ MANOMETER · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10889483110850), received between Jun 1, 2022 and Jun 18, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Decrease in Pressure1
  • Leak/Splash1
  • Manufacturing, Packaging or Shipping Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 18, 2025Break; Break

    IT WAS REPORTED, THE ENDOTRACHEAL TUBE WAS ATTACHED TO A BAGGED PATIENT WHEN THEY NOTICED PART OF THE CLEAR PLASTIC PART OF THE MANOMETER PIECE WAS BROKEN. IT WAS REPLACED WITH ANOTHER DEVICE; THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22286410 ↗
  • MalfunctionNov 17, 2023Leak/Splash; Leak/Splash

    THE BAG WAS LEAKING AIR. THE PATIENT IN THE PACU REQUIRED SUPPORT WITH BREATHING. THE CRNA REQUESTED A BAG VALVE MASK AND RECEIVED A 1L HYPERINFLATION SYSTEM + MANOMETER FROM VENTLAB (SUNMED). AFTER HOOKING UP THE OXYGEN THE CRNA ATTEMPTED VENTILATION AND NOTICED THE BAG LEAKING. ON INSPECTION, THE INFLATION BAG HAD SEVERAL SMALL CUTS THAT PREVENTED GOOD VENTILATION. A REPLACEMENT (SAME PRODUCT) WAS QUICKLY OBTAINED AND THE PATIENT WAS TREATED WITHOUT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18165397 ↗
  • MalfunctionOct 20, 2023

    THE PATIENT IN THE PACU REQUIRED SUPPORT WITH BREATHING. THE CRNA REQUESTED A BAG VALVE MASK AND RECEIVED A 1L HYPERINFLATION SYSTEM + MANOMETER FROM VENTLAB (SUNMED). AFTER HOOKING UP THE OXYGEN THE CRNA ATTEMPTED VENTILATION AND NOTICED THE BAG LEAKING. ON INSPECTION, THE INFLATION BAG HAD SEVERAL SMALL CUTS THAT PREVENTED GOOD VENTILATION. A REPLACEMENT (SAME PRODUCT) WAS QUICKLY OBTAINED AND THE PATIENT WAS TREATED WITHOUT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17975566 ↗
  • MalfunctionJun 1, 2022Decrease in Pressure

    THE COMMITTEE ON TECHNOLOGY FOR THE ANESTHESIA PATIENT SAFETY FOUNDATION RECEIVED A REPORT REGARDING OUR HYPERINFLATION SYSTEM. DESCRIBING THE DESIGN AS COUNTER-INTUITIVE AND THAT IT DOES NOT CONFORM WITH ISO STANDARD RECOMMENDATIONS. A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE. HOWEVER; THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. THE PROBLEM WAS REMEDIED BY QUICK OBSERVATION THAT TO CLOSE THE VALVE, THE VALVE MUST BE TURNED ANTI-CLOCKWISE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14570258 ↗
  • MalfunctionJun 1, 2022Manufacturing, Packaging or Shipping Problem

    THE COMMITTEE ON TECHNOLOGY FOR THE ANESTHESIA PATIENT SAFETY FOUNDATION RECEIVED A REPORT REGARDING OUR HYPERINFLATION SYSTEM DESCRIBING THE DESIGN AS COUNTER-INTUITIVE AND THAT IT DOES NOT CONFORM WITH ISO STANDARD RECOMMENDATIONS. A NEAR MISS OCCURRED WHERE POSITIVE PRESSURE WAS DESIRED AND THE VALVE WAS TURNED CLOCKWISE WITH THE EXPECTATION OF INCREASING PRESSURE, HOWEVER THIS LED TO DECREASING PRESSURE IN THE HYPERINFLATION SYSTEM. THE PROBLEM WAS REMEDIED BY QUICK OBSERVATION THAT TO CLOSE THE VALVE, THE VALVE MUST BE TURNED ANTI-CLOCKWISE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14569757 ↗

Other Resuscitator,Manual,Non Self-Inflating devices

All Resuscitator,Manual,Non Self-Inflating devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page