Atellica® VTR Syringes 11561350Recalls & reported problems
Sunmed, LLC · 3mL Vented, Luer Lock Syringe, 22g x 1" SafetyTip™, 50 IU (~16.7 IU/mL) Balanced Heparin, Vented Tip Cap · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10889483479483), received between Nov 14, 2024 and Mar 7, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Failure to Seal” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Seal3
- Fail-Safe Problem1
- Protective Measures Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Needle Stick/Puncture1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 7, 2025Failure to Seal; Failure to Seal
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 2028807-2025-00002 FOR THE FIRST REPORT. REFER TO 2028807-2025-00003 FOR THE SECOND REPORT. IT WAS REPORTED, THE SYRINGES WERE UNABLE TO ADEQUATELY RETAIN ARTERIAL BLOOD, DUE TO POOR SEALING OF THE PLUNGER WHICH RESULTED IN: BLOOD LOSS THROUGH THE PLUNGER; THE EFFICIENCY OF COLLECTIONS; DIFFICULTY IN CARRYING OUT ARTERIAL BLOOD SAMPLING, BLOOD LEAKING ONTO SURFACES AND INTO THE ENVIRONMENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21551976 ↗MalfunctionMar 7, 2025Failure to Seal; Failure to Seal
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 2028807-2025-00003 FOR THE SECOND REPORT, REFER TO 2028807-2025-00004 FOR THE THIRD REPORT. IT WAS REPORTED, THE SYRINGES WERE UNABLE TO ADEQUATELY RETAIN ARTERIAL BLOOD, DUE TO POOR SEALING OF THE PLUNGER WHICH RESULTED IN: BLOOD LOSS THROUGH THE PLUNGER; THE EFFICIENCY OF COLLECTIONS; DIFFICULTY IN CARRYING OUT ARTERIAL BLOOD SAMPLING, BLOOD LEAKING ONTO SURFACES AND INTO THE ENVIRONMENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21551955 ↗MalfunctionMar 7, 2025Failure to Seal; Failure to Seal
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 2028807-2025-00002 FOR THE FIRST REPORT, REFER TO 2028807-2025-00004 FOR THE THIRD REPORT. IT WAS REPORTED, THE SYRINGES WERE UNABLE TO ADEQUATELY RETAIN ARTERIAL BLOOD, DUE TO POOR SEALING OF THE PLUNGER WHICH RESULTED IN: BLOOD LOSS THROUGH THE PLUNGER; THE EFFICIENCY OF COLLECTIONS; DIFFICULTY IN CARRYING OUT ARTERIAL BLOOD SAMPLING, BLOOD LEAKING ONTO SURFACES AND INTO THE ENVIRONMENT. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21551954 ↗MalfunctionNov 14, 2024Fail-Safe Problem; Protective Measures Problem; Fail-Safe Problem; Protective Measures Problem
SUNMED HOLDINGS LLC., RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPERATE UNITS, INVOLVING THE SAME NURSE. THIS IS THE FIRST REPORT OF TWO REPORTS. REFER TO 2028807-2024-00052. IT WAS REPORTED A FAULT IN THE NEEDLE'S ANTI-PRICKING SAFETY SYSTEM. ALTHOUGH THE ANTI-PRICKING SYSTEM WAS ACTIVATED, THE TIP OF THE NEEDLE REMAINED EXPOSED, RESULTING IN THE ACCIDENTAL PRICKING OF A NURSE, NO PATENT WAS INVOLVED OR IMPACTED.
Patient problems: Needle Stick/Puncture; No Clinical Signs, Symptoms or Conditions
MAUDE report 20690001 ↗
Other Arterial Blood Sampling Kit devices
- MEDLINE ART995Medline Industries, Inc. · Arterial blood sampling kit4 recalls0 reports
- MEDLINE ART1070AMedline Industries, Inc. · Arterial blood sampling kit3 recalls0 reports
- MEDLINE ART1180Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART775BMedline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART890AMedline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- Centurion ART155ACenturion Medical Products Corporation · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART1160Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- OTHER P42453Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.