AirLife™ CSC100 (UDI 10889483598764)Recalls & reported problems

Airlife · Airlife Verso™ Airway Access Adapter · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889483598764), received between Oct 15, 2025 and Oct 15, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED: THE ANXIOUS INTUBATED PATIENT, IN THE INTENSIVE CARE UNIT (ICU), PULLED THE CLOSED SUCTION CATHETER (CSC) OFF FROM THE VERSO ADAPTER. ADDITIONALLY, THE DUCK VALVE, INSIDE THE VERSO ADAPTER CSC100, FELL DOWN INTO THE SWIVEL PART OF THE DEVICE, WHERE THE STAFF DISCOVERED IT. THE SYSTEM WAS CHANGED; THERE WAS NO DIRECT IMPACT ON THE PATIENT'S CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23305971 ↗

Other Circuit, Breathing (W Connector, Adaptor, Y Piece) devices