Medline Industries, Inc. DYNJ47948BRecalls & reported problems
Medline Industries, Inc. · MAJOR ORTHO PACK · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889942650231), received between Oct 24, 2018 and Nov 21, 2018 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Improper or Incorrect Procedure or Method1
Patient problems
- Burn(s)1
- Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 21, 2018Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT PATIENT SUSTAINED BURN (DISTAL TO THE SURGICAL SITE) FROM THE CAUTERY PENCIL THAT SIGNALED IN-USE ON GENERATOR DESPITE THE ROCKER-SWITCH ON THE CAUTERY PENCIL NOT ACTIVATED.
Patient problems: Burn(s)
MAUDE report 8096106 ↗MalfunctionOct 24, 2018Improper or Incorrect Procedure or Method
CAUTERY PENCIL FROM MAJOR ORTHO PACK WITH LOT NO 18HBK660 SIGNALED "IN USE" EVEN WHEN BUTTON WAS NOT PRESSED. CAUTERY PENCIL UNPLUGGED FROM ELECTRICAL UNIT. NEW PENCIL FROM OR STOCK PASSED TO FIELD. ALLEGED CUT/BURN TO PATIENT'S ARM.
Patient problems: Injury
MAUDE report 8000226 ↗
Other Orthopedic tray devices
- Medline Industries, Inc. DYNJ900718CMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ905330BMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ17452SMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ21826RMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. CDS780119VMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ902473LMedline Industries, Inc. · Orthopedic tray5 recalls0 reports
- Medline Industries, Inc. DYNJ906893DMedline Industries, Inc. · Orthopedic tray4 recalls2 reports
- Medline Industries, Inc. DYNJ909505AMedline Industries, Inc. · Orthopedic tray4 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page