Malibu 91-6117Recalls & reported problems

Seaspine Orthopedics Corporation · Thoracic Probe, Curved Square Tip · Class 1

All models of Malibu (17)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981082314), received between Dec 11, 2019 and Dec 11, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IN AN L5-S1 PLIF, THE DISTAL TIP OF THE 91-6116 (THORACIC PROBE, SQUARE TIP) BROKE OFF INTO THE S1 PEDICLE. DR. (B)(6) WAS MALLETING THE PROBE INTO THE PEDICLE WHEN IT BECAME STUCK. THE SURGEON WAS UNABLE TO REMOVE THE PROBE. UPON FORCEFULLY TWISTING THE INSTRUMENT TO RETRIEVE IT FROM THE PEDICLE, THE PROBE BROKE, LEAVING THE TIP IMPLANTED IN THE PEDICLE. THE SURGEON ELECTED TO LEAVE THE TIP IMPLANTED IN THE PATIENT, WHILE TAKING INTO CONSIDERATION THAT HE WOULD HAVE HAD TO DAMAGE THE PATIENT'S BONE AND/OR ANATOMY IN ORDER TO RECOVER THE TIP. DR. (B)(6) STATED THE PATIENT HAD UNUSUALLY HARD BONE AND THAT THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO PLANS TO REVISE. THE PATIENT IS HEALING AS EXPECTED, WITH NO COMPLICATIONS DUE TO THE PRODUCT MALFUNCTION AT THIS TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9451401 ↗
  • MalfunctionDec 11, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    IN AN L5-S1 PLIF, THE DISTAL TIP OF THE 91-6117 (THORACIC PROBE, CURVED SQUARE TIP) BROKE OFF INTO THE S1 PEDICLE. DR. (B)(6) WAS MALLETING THE PROBE INTO THE PEDICLE WHEN IT BECAME STUCK. THE SURGEON WAS UNABLE TO REMOVE THE PROBE. UPON FORCEFULLY TWISTING THE INSTRUMENT TO RETRIEVE IT FROM THE PEDICLE, THE PROBE BROKE, LEAVING THE TIP IMPLANTED IN THE PEDICLE. THE SURGEON ELECTED TO LEAVE THE TIP IMPLANTED IN THE PATIENT, WHILE TAKING INTO CONSIDERATION THAT HE WOULD HAVE HAD TO DAMAGE THE PATIENT'S BONE AND/OR ANATOMY IN ORDER TO RECOVER THE TIP. DR. (B)(6) STATED THE PATIENT HAD UNUSUALLY HARD BONE AND THAT THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO PLANS TO REVISE. THE PATIENT IS HEALING AS EXPECTED, WITH NO COMPLICATIONS DUE TO THE PRODUCT MALFUNCTION AT THIS TIME.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9451009 ↗