Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981160487), received between Jan 7, 2020 and Jan 7, 2020 — classified as 1 injury.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Device Embedded In Tissue or Plaque1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 7, 2020Fracture; Fracture

    ON (B)(6) 2019, A CABO ANTERIOR CERVICAL PLATE REVISION TOOK PLACE. THE ORIGINAL DATE OF SURGERY WAS (B)(6) 2019 WHERE THE PATIENT UNDERWENT SPINE SURGERY USING THE CABO ANTERIOR CERVICAL PLATE SYSTEM, CONTAINING A 3-LEVEL PLATE AND 8 SCREWS. THE CONSTRUCT ALSO CONSISTED OF 3 SHORELINE ACS INTERBODIES. A REVISION SURGERY TOOK PLACE ON (B)(6) 2019. IT WAS REPORTED THAT THE REVISION SURGERY WAS TO FUSE C7/T1 AND ALSO TO REVISE 2 SCREWS THAT LOOSENED AND 1 ADDITIONAL SCREW THAT HAD BACKED OUT DUE TO A CRACK IN THE CABO CERVICAL PLATE. IT WAS REPORTED THAT THE MOST INFERIOR LOCKING MECHANISM ON THE PLATE HAD BROKEN ALLOWING THE SCREW TO BACK OUT. THE REVISION SURGERY WAS TO REVISE AND REMOVE THE SCREWS AT THE LOWEST LEVEL AND ADD SHORELINE AT THE ADJACENT LEVEL TO THE PLATE. THE CRACKED CABO PLATE WAS LEFT IN THE PATIENT. THE REVISION SURGERY WAS COMPLETED SUCCESSFULLY AND THE PATIENT IS REPORTEDLY DOING WELL.

    Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient