Skip to content

Regatta Lateral System 36-90-3000Recalls & reported problems

Seaspine Orthopedics Corporation · Regatta Lateral Inserter · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10889981162306), received between Jun 28, 2023 and Jun 11, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2025Fracture; Fracture

    A PATIENT UNDERWENT LATERAL LUMBAR INTERBODY FUSION (LLIF) ON (B)(6) 2025 UTILIZING THE ORTHOFIX/SEASPINE WAVEFORM L INTERBODY SYSTEM. DURING INSERTION OF THE LATERAL INTERBODY, THE TIP OF THE REGATTA LATERAL INSERTER FRACTURED, AND THE INSTRUMENT FRAGMENT WAS LEFT IN THE IMPLANTED INTERBODY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SIGNIFICANT DELAY, AND THERE WERE NO REPORTED PATIENT INJURIES OR NEED FOR ADDITIONAL INTERVENTION.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22188716 ↗
  • MalfunctionMay 21, 2024Fracture; Fracture

    ON (B)(6) 2024, A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S REGATTA LATERAL SYSTEM. THE 36-90-3000 REGATTA LATERAL INSERTER FRACTURED INTRA-OPERATIVELY AND THE FRAGMENT WAS LEFT INSIDE THE PATIENT. THE CUSTOMER REPORTED THAT THE SURGERY WAS COMPLETED, AND NO ADDITIONAL TREATMENT WAS REQUIRED.

    Patient problems: Insufficient Information

    MAUDE report 19361804 ↗
  • MalfunctionJun 28, 2023Fracture; Fracture

    ON (B)(6) 2023 A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S REGATTA LATERAL SYSTEM. DURING IMPLANT INSERTION, THE REGATTA LATERAL INSERTER TIP FRACTURED AND WAS LEFT IN THE IMPLANT. THE IMPLANT WAS FULLY SEATED AND SECURED WITH BILATERAL PEDICLE SCREWS. NO PATIENT INJURY WAS REPORTED. THERE ARE NO PLANS FOR REVISION OR REMOVAL. NO RADIOGRAPHS WERE PROVIDED. SEASPINE WAS MADE AWARE OF THIS INCIDENT ON 09 JUN 2023.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17218840 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page