Skip to content

Admiral AP1-200030Recalls & reported problems

Seaspine Orthopedics Corporation · Cervical Plate, 2 Level, 30mm · Class 2 · Implant

All models of Admiral (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10889981238216), received between Feb 7, 2023 and Feb 7, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Premature Separation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Premature Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 7, 2023Premature Separation; Premature Separation

    A PATIENT UNDERWENT SPINAL SURGERY CONSISTING OF SEASPINE'S ADMIRAL PLATE SYSTEMS. THE INDEX SURGERY DATE WAS ON (B)(6) 2022. SEASPINE WAS MADE AWARE ON (B)(6) 2023 THAT A SCREW BACK OUT HAD OCCURRED. THE REPORTER NOTED THAT DURING THE INDEX SURGERY AN AP1-200030 CERVICAL PLATE, 2 LEVEL, 30MM WAS NOT ABLE TO CAM LOCK OVER THE SCREW TO FINAL LOCK. THEY ONLY HAD AN ISSUE ON ONE OF THE DISTAL ENDS OF THE PLATE. DURING A PHONE CALL ON (B)(6) 2023 WITH SEASPINE, THE REPORTER STATED THAT THE SURGEON DECIDED TO LEAVE THE PLATE IN PLACE AND COMPLETED THE INDEX SURGERY. HOWEVER, 4 MONTHS POSTOP THE SCREW BACKED OUT. THEN A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023 TO CORRECT THE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16313577 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page