Skip to content

Admiral AP2-100002Recalls & reported problems

Seaspine Orthopedics Corporation · Admiral ACP Screwdriver, Threaded · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10889981276775), received between May 12, 2025 and May 1, 2026 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 4 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Material Fragmentation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Foreign Body In Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 1, 2026Fracture

    ON 03 APR 2026, ORTHOFIX WAS MADE AWARE OF THE FOLLOWING EVENT UTILIZING THE SEASPINE ADMIRAL ACP SYSTEM. PER THE REPORTER, THE TIP OF THE SCREWDRIVER SHEARED OFF AND BECAME STUCK IN THE SCREW. THE FRACTURED TIP REMAINED IN THE SCREW AND COULD NOT BE REMOVED. THEY COULD ONLY PARTIALLY SLIDE THE LOCKING COVER OVER THE SCREWS. THE PATIENT WAS REPORTED TO BE UNINJURED.

    Patient problems: Foreign Body In Patient

    MAUDE report 25058661 ↗
  • MalfunctionMar 24, 2026Fracture

    ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT; HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 3 OF 4 UNITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24683270 ↗
  • MalfunctionMar 24, 2026Fracture

    ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 2 OF 4 UNITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24682948 ↗
  • MalfunctionMar 24, 2026Fracture

    ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 4 OF 4 UNITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24682864 ↗
  • MalfunctionMar 24, 2026Fracture

    ON (B)(6) 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 1 OF 4 UNITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24677356 ↗
  • MalfunctionMay 12, 2025Material Fragmentation; Material Fragmentation

    CUSTOMER REPORTS THAT DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, THE TIP OF THE AP2-100002 ADMIRAL ACP SCREWDRIVER SHEARED OFF. ALL PIECES WERE RETRIEVED; HOWEVER, THE ISSUE RESULTED IN A 30-MINUTE SURGICAL DELAY. NO ADDITIONAL MEDICAL OR SURGICAL TREATMENT WAS REQUIRED; THE SURGERY WAS ABLE TO BE COMPLETED AND THERE WERE NO ADVERSE PATIENT OUTCOMES. CUSTOMER STATES THE SURGEON USED THE PROPER TECHNIQUE WITH THE INSTRUMENTATION. MULTIPLE FOLLOW-UP ATTEMPTS WERE PERFORMED TO HAVE THE PRODUCT RETURNED FOR INVESTIGATION, HOWEVER, THE ADMIRAL ACP SCREWDRIVERS WERE NOT RETURNED AND ARE NOT AVAILABLE FOR EVALUATION. THIS OCCURRED WITH TWO SEPARATE ADMIRAL ACP SCREWDRIVERS. THIS MDR IS TO DOCUMENT THE EVENT AGAINST ONE LOT OF THE ADMIRAL ACP SCREWDRIVERS. SEE MDR 3012120772-2025-00031 FOR THE SECOND LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22007055 ↗
  • MalfunctionMay 12, 2025Material Fragmentation; Material Fragmentation

    CUSTOMER REPORTS THAT DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, THE TIP OF THE AP2-100002 ADMIRAL ACP SCREWDRIVER SHEARED OFF. ALL PIECES WERE RETRIEVED, HOWEVER, THE ISSUE RESULTED IN A 30-MINUTE SURGICAL DELAY. NO ADDITIONAL MEDICAL OR SURGICAL TREATMENT WAS REQUIRED, THE SURGERY WAS ABLE TO BE COMPLETED AND THERE WERE NO ADVERSE PATIENT OUTCOMES. CUSTOMER STATES THE SURGEON USED THE PROPER TECHNIQUE WITH THE INSTRUMENTATION. MULTIPLE FOLLOW-UP ATTEMPTS WERE PERFORMED TO HAVE THE PRODUCT RETURNED FOR INVESTIGATION, HOWEVER, THE ADMIRAL ACP SCREWDRIVERS WERE NOT RETURNED AND ARE NOT AVAILABLE FOR EVALUATION. THIS OCCURRED WITH TWO SEPARATE ADMIRAL ACP SCREWDRIVERS. THIS MDR IS TO DOCUMENT THE EVENT AGAINST THE SECOND LOT OF THE ADMIRAL ACP SCREWDRIVERS. SEE MDR 3012120772-2025-00030 FOR THE FIRST LOT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22007043 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page