Admiral AP2-100002Recalls & reported problems
Seaspine Orthopedics Corporation · Admiral ACP Screwdriver, Threaded · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10889981276775), received between May 12, 2025 and May 1, 2026 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Fracture” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Material Fragmentation2
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Foreign Body In Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2026Fracture
ON 03 APR 2026, ORTHOFIX WAS MADE AWARE OF THE FOLLOWING EVENT UTILIZING THE SEASPINE ADMIRAL ACP SYSTEM. PER THE REPORTER, THE TIP OF THE SCREWDRIVER SHEARED OFF AND BECAME STUCK IN THE SCREW. THE FRACTURED TIP REMAINED IN THE SCREW AND COULD NOT BE REMOVED. THEY COULD ONLY PARTIALLY SLIDE THE LOCKING COVER OVER THE SCREWS. THE PATIENT WAS REPORTED TO BE UNINJURED.
Patient problems: Foreign Body In Patient
MAUDE report 25058661 ↗MalfunctionMar 24, 2026Fracture
ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT; HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 3 OF 4 UNITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24683270 ↗MalfunctionMar 24, 2026Fracture
ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 2 OF 4 UNITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24682948 ↗MalfunctionMar 24, 2026Fracture
ON 25 FEB 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 4 OF 4 UNITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24682864 ↗MalfunctionMar 24, 2026Fracture
ON (B)(6) 2026, ORTHOFIX WAS MADE AWARE OF MULTIPLE ADMIRAL DRIVER TIP FRACTURES. IT WAS REPORTED THAT THE FRACTURED TIPS BECAME LODGED IN THE SCREW SHANK AND WERE UNABLE TO BE RETRIEVED WITHOUT REMOVING THE ENTIRE SCREW. THIS RESULTED IN A SURGICAL DELAY OF GREATER THAN 30 MINUTES. NO PATIENT INJURIES WERE REPORTED. QUALITY COMPLIANCE MADE MULTIPLE ATTEMPTS TO OBTAIN FURTHER INFORMATION REGARDING THE EVENT, HOWEVER, CUSTOMER WAS UNRESPONSIVE. IT IS UNKNOWN IF THE ENTIRE SCREW WAS REPLACED OR IF THE SCREW WAS LEFT IN-SITU WITH THE UNRETRIEVED FRAGMENT. THIS REPORT IS FOR 1 OF 4 UNITS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24677356 ↗MalfunctionMay 12, 2025Material Fragmentation; Material Fragmentation
CUSTOMER REPORTS THAT DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, THE TIP OF THE AP2-100002 ADMIRAL ACP SCREWDRIVER SHEARED OFF. ALL PIECES WERE RETRIEVED; HOWEVER, THE ISSUE RESULTED IN A 30-MINUTE SURGICAL DELAY. NO ADDITIONAL MEDICAL OR SURGICAL TREATMENT WAS REQUIRED; THE SURGERY WAS ABLE TO BE COMPLETED AND THERE WERE NO ADVERSE PATIENT OUTCOMES. CUSTOMER STATES THE SURGEON USED THE PROPER TECHNIQUE WITH THE INSTRUMENTATION. MULTIPLE FOLLOW-UP ATTEMPTS WERE PERFORMED TO HAVE THE PRODUCT RETURNED FOR INVESTIGATION, HOWEVER, THE ADMIRAL ACP SCREWDRIVERS WERE NOT RETURNED AND ARE NOT AVAILABLE FOR EVALUATION. THIS OCCURRED WITH TWO SEPARATE ADMIRAL ACP SCREWDRIVERS. THIS MDR IS TO DOCUMENT THE EVENT AGAINST ONE LOT OF THE ADMIRAL ACP SCREWDRIVERS. SEE MDR 3012120772-2025-00031 FOR THE SECOND LOT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22007055 ↗MalfunctionMay 12, 2025Material Fragmentation; Material Fragmentation
CUSTOMER REPORTS THAT DURING A SURGERY THAT TOOK PLACE ON (B)(6) 2025, THE TIP OF THE AP2-100002 ADMIRAL ACP SCREWDRIVER SHEARED OFF. ALL PIECES WERE RETRIEVED, HOWEVER, THE ISSUE RESULTED IN A 30-MINUTE SURGICAL DELAY. NO ADDITIONAL MEDICAL OR SURGICAL TREATMENT WAS REQUIRED, THE SURGERY WAS ABLE TO BE COMPLETED AND THERE WERE NO ADVERSE PATIENT OUTCOMES. CUSTOMER STATES THE SURGEON USED THE PROPER TECHNIQUE WITH THE INSTRUMENTATION. MULTIPLE FOLLOW-UP ATTEMPTS WERE PERFORMED TO HAVE THE PRODUCT RETURNED FOR INVESTIGATION, HOWEVER, THE ADMIRAL ACP SCREWDRIVERS WERE NOT RETURNED AND ARE NOT AVAILABLE FOR EVALUATION. THIS OCCURRED WITH TWO SEPARATE ADMIRAL ACP SCREWDRIVERS. THIS MDR IS TO DOCUMENT THE EVENT AGAINST THE SECOND LOT OF THE ADMIRAL ACP SCREWDRIVERS. SEE MDR 3012120772-2025-00030 FOR THE FIRST LOT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22007043 ↗
Other Appliance, fixation, spinal intervertebral body devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Elegance® Anterior Cervical Plate 53430353Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430108Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430114Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430240Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430236Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page