Skip to content

Admiral AP1-704014Recalls & reported problems

Seaspine Orthopedics Corporation · 4.0mm Variable Angle Screw, Self-Tapping, 14mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10889981278823), received between Jun 2, 2025 and Jun 2, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • Insufficient Information2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 2, 2025Fracture; Fracture

    ON 05 MAY 2025, SEASPINE/ORTHOFIX WAS MADE AWARE OF A REVISION SURGERY DUE TO A SCREW THAT BACKED OUT OF AN ADMIRAL CERVICAL PLATE CONSTRUCT. THE PROVIDED X-RAY CORROBORATES THE REPORT. TWO SCREWS AND THE PLATE'S INFERIOR LOCKING COVER WERE REMOVED. THE SURGEON CHOSE TO LEAVE THE PLATE AND REMAINING SCREWS IN-SITU. NO PATIENT INJURY WAS REPORTED. THIS REPORT IS FOR 1 OF 2 AP1-704014 4.0MM VARIABLE ANGLE SCREWS, SELF-TAPPING, 14MM. THE AP1-310045 CERVICAL PLATE, 3 LEVEL, 45MM WAS REPORTED ON 3012120772-2025-00034.

    Patient problems: Insufficient Information

    MAUDE report 22127702 ↗
  • InjuryJun 2, 2025Fracture; Fracture

    ON (B)(6) 2025, SEASPINE/ORTHOFIX WAS MADE AWARE OF A REVISION SURGERY DUE TO A SCREW THAT BACKED OUT OF AN ADMIRAL CERVICAL PLATE CONSTRUCT. THE PROVIDED X-RAY CORROBORATES THE REPORT. TWO SCREWS AND THE PLATE'S INFERIOR LOCKING COVER WERE REMOVED. THE SURGEON CHOSE TO LEAVE THE PLATE AND REMAINING SCREWS IN-SITU. NO PATIENT INJURY WAS REPORTED. THIS REPORT IS FOR 2 OF 2 AP1-704014 4.0MM VARIABLE ANGLE SCREWS, SELF-TAPPING, 14MM. THE AP1-310045 CERVICAL PLATE, 3 LEVEL, 45MM WAS REPORTED ON 3012120772-2025-00034.

    Patient problems: Insufficient Information

    MAUDE report 22127688 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates