HUDSON RCI IPN913729Recalls & reported problems

Teleflex Incorporated · Confort Flo Humidification System with · Class 2

All models of HUDSON RCI (152)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 14026704659354), received between Jan 26, 2021 and Jan 26, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Deformation Due to Compressive Stress” (1 report).

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 13, 2019Class ITerminated

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

    Z-1115-2020 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Deformation Due to Compressive Stress1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 26, 2021Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    IT WAS REPORTED "THAT THE PLASTIC TUBING FROM THE CONCHA COLUMN WILL KINK OFF DUE TO THE FLIMSINESS". UNKNOWN PATIENT CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11229671 ↗

Other HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) devices

All HUMIDIFIER, RESPIRATORY GAS, (DIRECT PATIENT INTERFACE) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page