Skip to content

HUDSON RCI IPN913731Recalls & reported problems

Teleflex Incorporated · COMFORT FLO · Class 2

All models of HUDSON RCI (152)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Dec 13, 2019Class ITerminated

    There is the potential for water to flood the column and enter the circuit under circumstances where an abnormal pressure differential is created between the water bottle and the column during high flow oxygen therapy.

    Z-1113-2020 · Root cause: Device Design · Recall details · FDA record ↗

Other Humidifier nebulizer kit devices

All Humidifier nebulizer kit devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page