LMA ATOMIZATION IPN915303Recalls & reported problems

Teleflex Incorporated · MAD Nasal with 3mL Syringe · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
18
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 14026704709813), received between Jan 7, 2025 and Nov 3, 2025 — classified as 18 malfunctions.

The most frequently reported specific device problem is “Material Discolored” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 6 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Discolored6
  • Complete Blockage3
  • Product Quality Problem3
  • Contamination /Decontamination Problem2
  • Device Markings/Labelling Problem2
  • Lack of Maintenance Documentation or Guidelines2
  • Delivered as Unsterile Product1
  • Inadequate Instructions for Healthcare Professional1

Patient problems

  • No Clinical Signs, Symptoms or Conditions18

Recent adverse event reports

The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 3, 2025Inadequate Instructions for Healthcare Professional; Inadequate Instructions for Healthcare Professional

    IT WAS REPORTED THAT " MISSING MODEL AND LOT DATA ON THE PACKAGING. THE ISSUE WAS IDENTIFIED AT INCOMING INSPECTION."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23449117 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    IT WAS REPORTED THAT " STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION." ASSOCIATED COMPLAINTS: "300 3898360-2025-00355, 3003898360-2025-00347 AND 3003898360-2025-00354.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898190 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    IT WAS REPORTED THAT " STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION." ASSOCIATED COMPLAINTS: "300 3898360-2025-00355, 3003898360-2025-00347."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898175 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    FAILURE MODE "FOREIGN MATERIAL - STAIN" COULD BE CONFIRMED WITH PICTURE ATTACHED. HOWEVER IT IS NECESSARY TO RECEIVE THE PHYSICAL SAMPLE TO PERFORM A PROPER INVESTIGATION, DETERMINATE THE ROOT CAUSE & CORRECTIVE ACTIONS. IF THE COMPLAINT SAMPLES BECOME AVAILABLE AT A LATER DATE, THIS COMPLAINT WILL BE UPDATED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898073 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    IT WAS REPORTED THAT " STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION." ASSOCIATED COMPLAINTS: "300 3898360-2025-00355, 3003898360-2025-00347 AND 3003898360-2025-00354.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898044 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    IT WAS REPORTED THAT " STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION." ASSOCIATED COMPLAINTS: "3003898360-2025-00355, 3003898360-2025-00347 AND 3003898360-2025-00354.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898026 ↗
  • MalfunctionApr 24, 2025Material Discolored; Material Discolored

    IT WAS REPORTED THAT " STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION." ASSOCIATED COMPLAINTS: "300 3898360-2025-00355 AND 3003898360-2025-00354."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21898014 ↗
  • MalfunctionMar 6, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT "MISSING DATA ON THE PRIMARY PACKAGING. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION. THERE WAS NO PATIENT INVOLVED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534489 ↗
  • MalfunctionMar 6, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT "FOREIGN MATERIAL IN THE PRIMARY PACKAGING. THE ISSUE WAS IDENTIFIED DURING INCOMING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT.".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534361 ↗
  • MalfunctionMar 6, 2025Contamination /Decontamination Problem; Contamination /Decontamination Problem

    IT WAS REPORTED THAT "STAIN ON THE NASAL ADAPTOR. THE ISSUE WAS IDENTIFIED ON INCOMING INSPECTION. THERE WAS NO PATIENT INVOLVEMENT."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534097 ↗
  • MalfunctionMar 5, 2025Lack of Maintenance Documentation or Guidelines; Lack of Maintenance Documentation or Guidelines

    IT WAS REPORTED THAT "MISSING DATA ON THE LABEL. ISSUE WAS IDENTIFIED PRIOR TO USE. ASSOCIATED COMPLAINTS 3003898360-2025-00206, 3003898360-2025-00242 AND 3003898360-2025-00243.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21524186 ↗
  • MalfunctionMar 4, 2025Lack of Maintenance Documentation or Guidelines; Lack of Maintenance Documentation or Guidelines

    IT WAS REPORTED THAT "MISSING DATA ON THE LABEL. ISSUE WAS IDENTIFIED PRIOR TO USE. ASSOCIATED COMPLAINTS 3003898360-2025-00206, 300 3898360-2025-00242 .

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514793 ↗
  • MalfunctionMar 4, 2025Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT "MISSING DATA ON THE LABEL".

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21514093 ↗
  • MalfunctionMar 4, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT "STAIN ON THE PRIMARY PACKAGING. THE ISSUE WAS IDENTIFIED AT INCOMING INSPECTION /INVENTORY REVIEW. THERE WAS NO PATIENT INVOLVED."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21512840 ↗
  • MalfunctionMar 4, 2025Unsealed Device Packaging; Unsealed Device Packaging

    IT WAS REPORTED THAT: "INCOMPLETE SEALING OF THE PACKAGING. THE ISSUE WAS IDENTIFIED AT INCOMING INSPECTION /INVENTORY REVIEW."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21512826 ↗
  • MalfunctionJan 7, 2025Complete Blockage; Product Quality Problem; Complete Blockage; Product Quality Problem

    IT WAS REPORTED THAT: "CONE OF THE NASAL CONNECTOR COMPLETELY CLOSED. ALSO, OTHERS FROM VE WITH THE SAME PRODUCTION ERROR. THERE WAS NO PATIENT INVOLVED" ASSOCIATED COMPLAINTS 3003898360-2025-00028, 3003898360-2025-00029 AND (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21089142 ↗
  • MalfunctionJan 7, 2025Complete Blockage; Product Quality Problem; Complete Blockage; Product Quality Problem

    IT WAS REPORTED THAT: "CONE OF THE NASAL CONNECTOR COMPLETELY CLOSED. ALSO, OTHERS FROM VE WITH THE SAME PRODUCTION ERROR. THERE WAS NO PATIENT INVOLVED." ASSOCIATED COMPLAINTS 3003898360-2025-00028, 3003898360-2025-00029 AND (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21089121 ↗
  • MalfunctionJan 7, 2025Complete Blockage; Product Quality Problem; Complete Blockage; Product Quality Problem

    IT WAS REPORTED THAT: "CONE OF THE NASAL CONNECTOR COMPLETELY CLOSED. ALSO OTHERS FROM VE WITH THE SAME PRODUCTION ERROR." ASSOCIATED COMPLAINTS 3003898360-2025-00028, .AND (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21089103 ↗

Other NEBULIZER, MEDICINAL, NON-VENTILATORY (ATOMIZER) devices

All NEBULIZER, MEDICINAL, NON-VENTILATORY (ATOMIZER) devices →