LMA IPN924713Recalls & reported problems
Teleflex Incorporated · ARROW AutoSelector Disposable Pain Control Pump · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14026704743541), received between Aug 2, 2022 and Jul 18, 2023 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Excess Flow or Over-Infusion” (2 reports).
16 other recalls cover devices cleared under the same FDA submission (K090300); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Excess Flow or Over-Infusion2
- Disconnection1
- Leak/Splash1
Patient problems
- Insufficient Information1
- Lethargy1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJul 18, 2023Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, "THE AUTOFUSER WAS SET TO 8ML PER HOUR AND THE AUTOFUSER EMPTIED COMPLETELY IN 24 HOURS. THE PATIENT HAD A WOUND VAC ON". NO PATIENT HARM OR INJURY. THE PATIENT STATUS IS REPORTED AS "FINE". ADDITIONAL INFORMATION RECEIVED STATED THAT THE "PATIENT WAS SOMNOLENT AFTER 24 HOURS OF INFUSION (500ML ROPIVACAINE INFUSED FROM AUTOFUSER IN THAT TIME PERIOD). 20% LIPIDS WERE ADMINISTERED (350ML) AND PATIENT'S MENTAL STATUS RETURNED TO BASELINE."
Patient problems: Lethargy
MAUDE report 17342613 ↗InjuryJul 18, 2023Excess Flow or Over-Infusion; Excess Flow or Over-Infusion
IT WAS REPORTED THAT WHILE IN USE ON A PATIENT, "THE AUTOFUSER WAS SET TO 10ML PER HOUR AND THE AUTOFUSER EMPTIED COMPLETELY IN 24 HOURS. THE PATIENT HAD A WOUND VAC ON". ADDITIONAL INFORMATION RECEIVED STATED THAT "THE PATIENT WAS LETHARGIC AND DIFFICULT TO AROUSE. LIPIDS WERE INFUSED TO COUNTERACT THE ROPIVACAINE AND THE PATIENT SEEMED TO RESPOND TO THIS. THE ISSUE WAS RESOLVED BY DISCONTINUING THE AUTOFUSER WITH ROPIVACAINE".
Patient problems: Insufficient Information
MAUDE report 17342416 ↗MalfunctionAug 2, 2022Disconnection; Leak/Splash; Disconnection; Leak/Splash
COMPLAINT REPORTED THAT: "PUMP WAS HOOKED UP TO THE PATIENT AND WHILE PATIENT WAS STILL IN RECOVERY THE LINE BROKE AT THE CONNECTION POINT AND STARTED TO LEAK". NO PATIENT HARM REPORTED. NO REPORT OF MEDICAL INTERVENTION. PATIENT CONDITION REPORTED AS FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15145197 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090300. One submission often covers many models, so check the product and lot details in each record.
- Z-0940-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 1-14ML/HR CPNB, Product code MVBXL- CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0950-2019 · Oct 2, 2018 · TerminatedAF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0951-2019 · Oct 2, 2018 · TerminatedAF 550x8ML/H 2ML 15LO 8.7" DUAL CATHS, 4, Product code MT8015XLYK10 CP, Arrow AutoFuser Disposable Pain Control Pump
- Z-0941-2019 · Oct 2, 2018 · TerminatedAF w/AS 550x 0.5 7 ml/hr, 5ml 30 LO CPNB, Product code MVAX30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0955-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 1-14ML/HR 2ML-30LO CPNB, Product code MVBT30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0946-2019 · Oct 2, 2018 · TerminatedAF 550 x 5ML/H 6" DUAL CATHS, 4.5", Product code MC0050XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump
- Z-0942-2019 · Oct 2, 2018 · TerminatedAF550x5ML/H 2ML 60LO 6" CATH, 4.5" IN, Product code MT5060XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump
- Z-0954-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 1-14ML/HR 2ML-15LO CPNB, Product code MVBT15XL-CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0943-2019 · Oct 2, 2018 · TerminatedAF550x5ML/H 2ML 60LO 6" DUAL CATHS, 4.5", Product MT5060XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump
- Z-0952-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 0.5-7 ml,w/Y site,DL CPNB, Product code MVAXL2Y CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0953-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 1-14 ml/hr, 5ml30 LOCPNB, Product code MVBX30XLCPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0956-2019 · Oct 2, 2018 · TerminatedAF w/AS 550 x 1-14ML/HR 5ML-60LO CPNB, Product code MVBX60XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0945-2019 · Oct 2, 2018 · TerminatedAF 550 x 5ML/H 6" CATH, 4.5" IN, Product code MC0050XLSK5 CP, Arrow AutoFuser Disposable Pain Control Pump
- Z-0948-2019 · Oct 2, 2018 · TerminatedAF 550 x 5ML/HR 2ML 60LO CPNB, Product code MT5060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
- Z-0947-2019 · Oct 2, 2018 · TerminatedAF 550 x 5ML/H FH DL 30hole SAT C,4"IN, Product code MC0050XLYK2 CP, Arrow AutoFuser Disposable Pain Control Pump
…and 1 more.
Other Pump, infusion, elastomeric devices
- EASYPUMP 4540018-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC313 reports
- Halyard 13568Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC86 reports
- ON-Q* Pain Relief System with SELECT-A-FLOW 13472Avanos Medical, Inc. · Pump, infusion, elastomeric73 reports
- Halyard 13472Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC68 reports
- EASYPUMP 4540008-02B. Braun Medical Inc. · PUMP, INFUSION, ELASTOMERIC54 reports
- DOSI-FUSER L25915-500D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- DOSI-FUSER L25915-600D1-USALeventon Sau · Pump, Infusion, Elastomeric1 recall0 reports
- Halyard 13561Avanos Medical, Inc. · PUMP, INFUSION, ELASTOMERIC42 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page