NanoBone® SBX Putty 200046Recalls & reported problems
Biocomposites GmbH · Synthetic biodegradable bone graft substitute consisting of NanoBone® | granulate (non-sintered nanocrystalline hydroxyapatite in a porous silica gel matrix) embedded in an aqueous gel · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 14260418440468), received between Mar 7, 2023 and Sep 6, 2023 — classified as 12 injuries.
The most frequently reported specific device problem is “Migration” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Migration6
- Adverse Event Without Identified Device or Use Problem5
- Migration or Expulsion of Device4
Patient problems
- Pain5
- Hematoma4
- Post Operative Wound Infection3
- Iritis1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
InjurySep 6, 2023Migration; Migration
THIS EMDR WAS ORIGINALLY SUBMITTED IN AUGUST 2020 WITH A FOLLOW-UP IN SEPTEMBER 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD INTO FDA'S EMDR SYSTEM CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: A REPORT OF POSSIBLE MIGRATION OF A PORTION OF THE NANOBONE SBX PUTTY AND /OR INFECTION OF THE SURGICAL SITE WAS RECEIVED FROM A PHYSICIAN'S ASSISTANT. PATIENT WAS OPERATED ON FOR ONE LEVEL LUMBAR SPINAL FUSION. PATIENT EXPERIENCED POST-OPERATIVE EDEMA AND EXCESSIVE DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. INCISION WAS WASHED OUT, THEN CLOSED AND DRAINAGE AND EDEMA RESOLVED.
Patient problems: Hematoma
MAUDE report 17695622 ↗InjurySep 6, 2023Migration; Migration
THIS EMDR WAS ORIGINALLY SUBMITTED IN AUGUST 2020 WITH A FOLLOW-UP IN SEPTEMBER 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD INTO FDA'S EMDR SYSTEM CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: A REPORT OF POSSIBLE MIGRATION OF A PORTION OF THE NANOBONE SBX PUTTY AND /OR INFECTION OF THE SURGICAL SITE WAS RECEIVED FROM A PHYSICIAN'S ASSISTANT. PATIENT WAS OPERATED ON FOR ONE LEVEL LUMBAR SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE POST-OP EDEMA AND PAIN AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. INCISION WAS WASHED OUT, THEN CLOSED AND EDEMA AND PAIN RESOLVED.
Patient problems: Hematoma
MAUDE report 17695597 ↗InjurySep 6, 2023Migration; Migration
THIS EMDR ALONG WITH A FOLLOW-UP WERE ORIGINALLY SUBMITTED IN AUGUST 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD INTO FDA'S EMDR SYSTEM CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: A REPORT OF POSSIBLE MIGRATION OF A PORTION OF THE NANOBONE SBX PUTTY AND /OR INFECTION OF THE SURGICAL SITE WAS RECEIVED FROM A PHYSICIAN'S ASSISTANT. PATIENT WAS OPERATED ON FOR ONE LEVEL POSTERIOR CERVICAL SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE WOUND DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. THIS WAS SUPPORTED BY RADIOGRAPHS AND PHOTOGRAPHS OF THE OPENED INCISION. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. INCISION WAS WASHED OUT, THEN CLOSED AND WOUND DRAINAGE RESOLVED.
Patient problems: Post Operative Wound Infection
MAUDE report 17695462 ↗InjurySep 6, 2023Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration
THIS EMDR WAS ORIGINALLY SUBMITTED IN AUGUST 2020 WITH FOLLOW-UP IN SEPTEMBER 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD INTO FDA'S EMDR SYSTEM CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: A REPORT OF POSSIBLE MIGRATION OF A PORTION OF THE NANOBONE SBX PUTTY AND /OR INFECTION OF THE SURGICAL SITE WAS RECEIVED FROM A PHYSICIAN'S ASSISTANT FOR AN UNKNOWN SURGICAL PROCEDURE/SURGICAL INTERVENTION. PATIENT WAS OPERATED ON FOR ONE LEVEL POSTERIOR CERVICAL SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE WOUND DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. INCISION WAS WASHED OUT, THEN CLOSED AND WOUND DRAINAGE RESOLVED.
Patient problems: Post Operative Wound Infection
MAUDE report 17690646 ↗InjurySep 6, 2023Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration
THIS EMDR WAS ORIGINALLY SUBMITTED IN JULY 2020 WITH FOLLOW-UP IN SEPTEMBER 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD INTO FDA'S EMDR SYSTEM CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: A REPORT OF POSSIBLE MIGRATION OF A PORTION OF THE NANOBONE SBX PUTTY AND/OR INFECTION OF THE SURGICAL SITE WAS RECEIVED BUT DETAILS OF THE EVENT WERE NOT PROVIDED. IT WAS LATER LEARNED THAT THE PATIENT WAS OPERATED ON FOR THREE-LEVEL POSTERIOR CERVICAL SPINAL FUSION, AND THAT THE PATIENT EXPERIENCED POST-OPERATIVE EDEMA AND PAIN AT SURGERY SITE. AT THE FIVE-WEEK FOLLOW-UP VISIT, THE INCISION WAS REOPENED AND WASHED OUT. ADDITIONAL BONE GRAFT WAS PLACED. THE INCISION WAS CLOSED AND THE EDEMA AND PAIN RESOLVED.
Patient problems: Hematoma
MAUDE report 17689716 ↗InjurySep 6, 2023Adverse Event Without Identified Device or Use Problem; Migration; Adverse Event Without Identified Device or Use Problem; Migration
THIS EMDR WAS ORIGINALLY SUBMITTED IN JULY 2020 WITH FOLLOW-UP IN SEPTEMBER 2020. IT WAS LATER DISCOVERED THAT THOSE REPORTS APPARENTLY DID NOT UPLOAD CORRECTLY. ACCORDINGLY, THEY ARE BEING CONSOLIDATED AND SUBMITTED AGAIN HERE TO MAKE SURE THE REPORTED INFORMATION IS IN THE EMDR SYSTEM AS REQUIRED. SPECIFICALLY: PATIENT RECEIVED NANOBONE SBX PUTTY IN THE POSTEROLATERAL INTERTRANSVERSE PROCESS SPACE FOR SINGLE LEVEL SPINAL FUSION. AT FIRST FOLLOW-UP VISIT, PATIENT EXHIBITED MIGRATION OF A PORTION OF THE GRAFT TO BENEATH THE SKIN OVER THE FUSION SITE ON RADIOGRAPHS. PATIENT ALSO EXHIBITED POST-OPERATIVE EDEMA AND EXPERIENCED PAIN AT THE SURGICAL SITE. AT THREE-WEEK FOLLOW-UP VISIT, THE SURGICAL INCISION WAS REOPENED, HEMATOMA WAS DRAINED, AND INCISION WAS WASHED OUT. THERE APPEARED TO BE GRANULES OF NANOBONE SBX PUTTY IN THE HEMATOMA. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. THE INCISION WAS THEN CLOSED, AND EDEMA AND PAIN RESOLVED. PATIENT WENT ON TO SUCCESSFUL FUSION.
Patient problems: Hematoma; Post Operative Wound Infection
MAUDE report 17689705 ↗InjuryMar 13, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
PATIENT WAS OPERATED ON FOR ONE LEVEL POSTERIOR CERVICAL SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE WOUND DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. INCISION WAS WASHED OUT, THEN CLOSED AND WOUND DRAINAGE RESOLVED.
Patient problems: Pain
MAUDE report 16528055 ↗InjuryMar 13, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
PATIENT WAS OPERATED ON FOR ONE LEVEL LUMBAR SPINAL FUSION. PATIENT EXPERIENCED POST-OPERATIVE EDEMA AND EXCESSIVE DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. INCISION WAS WASHED OUT, THEN CLOSED AND DRAINAGE AND EDEMA RESOLVED.
Patient problems: Pain
MAUDE report 16528000 ↗InjuryMar 13, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
PATIENT WAS OPERATED ON FOR ONE LEVEL LUMBAR SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE POST-OP EDEMA AND PAIN AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. INCISION WAS WASHED OUT, THEN CLOSED AND EDEMA AND PAIN RESOLVED.
Patient problems: Pain
MAUDE report 16527990 ↗InjuryMar 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS OPERATED ON FOR THREE LEVEL POSTERIOR CERVICAL SPINAL FUSION. PATIENT EXPERIENCED POST-OPERATIVE EDEMA AND PAIN AT SURGERY SITE. AT FIVE WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND WASHED OUT. ADDITIONAL BONE GRAFT WAS PLACED. INCISION WAS CLOSED AND EDEMA AND PAIN RESOLVED.
Patient problems: Pain
MAUDE report 16527931 ↗InjuryMar 7, 2023Migration or Expulsion of Device; Migration or Expulsion of Device
PATIENT WAS OPERATED ON FOR ONE LEVEL POSTERIOR CERVICAL SPINAL FUSION. PATIENT EXPERIENCED EXCESSIVE WOUND DRAINAGE AT SURGERY SITE. AT TWO WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND SOME NANOBONE BONE GRAFT APPEARED TO HAVE MIGRATED FROM THE IMPLANTATION SITE TO BENEATH THE SKIN. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. INCISION WAS WASHED OUT, THEN CLOSED AND WOUND DRAINAGE RESOLVED.
Patient problems: Pain
MAUDE report 16494912 ↗InjuryMar 7, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
PATIENT WAS OPERATED ON FOR SINGLE LEVEL LUMBAR SPINAL FUSION. PATIENT EXPERIENCED POST-OPERATIVE EDEMA AND PAIN AT SURGERY SITE. AT THREE WEEK FOLLOW-UP VISIT, INCISION WAS REOPENED AND WASHED OUT. SPECIMEN TAKEN FROM SURGERY SITE SHOWED INFECTION, ORGANISM NOT IDENTIFIED. INCISION WAS CLOSED AND EDEMA AND PAIN RESOLVED. PATIENT WENT ON TO SUCCESSFUL FUSION.
Patient problems: Iritis
MAUDE report 16494784 ↗
FDA clearance and classification
The FDA regulates product code MQV (filler, bone void, calcium compound) under 21 CFR 888.3045 as a Class II device. 21 CFR 888.3045 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NanoBone® SBX Putty 200046
Has the NanoBone® SBX Putty 200046 been recalled?
No FDA recall record in our data names this exact device identifier (DI 14260418440468), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NanoBone® SBX Putty 200046?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Mar 7, 2023 and Sep 6, 2023. In those reports, the most common specific device problems are “Migration” (6) and “Migration or Expulsion of Device” (4); the most common patient problems are “Pain” (5), “Hematoma” (4) and “Post Operative Wound Infection” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NanoBone® SBX Putty 200046?
None of the reports are classified as deaths, 12 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NanoBone® SBX Putty 200046 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the NanoBone® SBX Putty 200046, and is it still sold?
The FDA's device database lists Biocomposites GmbH as the company responsible for it (brand NanoBone® SBX Putty, model 200046). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Filler, Bone Void, Calcium Compound devices
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44150Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND3 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42275Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls1 report
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T44145Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42200Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T42280Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND2 recalls0 reports
- MAGNIFUSE® Bone Graft 7509014Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- GRAFTON®and GRAFTON PLUS®Demineralized Bone Matrix (DBM) T43102Medtronic Sofamor Danek, Inc. · FILLER, BONE VOID, CALCIUM COMPOUND1 recall2 reports
- HydroSet XT 897005Stryker Corporation · FILLER, BONE VOID, CALCIUM COMPOUND1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.