PK7300 N3209000Recalls & reported problems
Beckman Coulter, Inc. · PK7300 · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 14987666535721), received between Feb 2, 2018 and Aug 27, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Defective Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Defective Component1
- Incorrect, Inadequate or Imprecise Result or Readings1
- Sparking1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Patient Involvement1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 27, 2021Defective Component; Defective Component
BECKMAN FIELD SERVICE ENGINEER (FSE) PERFORMED A PREVENTIVE MAINTENANCE (PM), THE FSE HAD POWERED UP THE SYSTEM AFTER REPLACING PM PARTS AND WENT OUT FOR A BREAK WHILE SYSTEM INITIALIZED. THE CUSTOMER CALLED FSE AND REPORTED INSTRUMENT SPARKING AND OBSERVED FLAMES. AN OPEN FLAME AND SPARK OCCURRED FROM THE BACK LEFT SIDE OF THE ANALYZER. NO ONE WAS INJURED BUT FIRE EXTINGUISHER WAS DISCHARGED BY THE CUSTOMER. BUILDING WAS EVACUATED AND FIRE DEPARTMENT WAS CALLED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12379069 ↗MalfunctionApr 12, 2021Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
THE CUSTOMER REPORTED ERRONEOUS/QUESTIONED RESULTS WITH O FLAGS ON THEIR PK7300 ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY OR IMPACT TO PATIENT CARE ASSOCIATED WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE ANY PATIENT DATA OR DEMOGRAPHICS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11653680 ↗MalfunctionFeb 2, 2018Sparking; Sparking
THE CUSTOMER REPORTED OBTAINING MULTIPLE ERROR MESSAGES ON THE LABORATORY'S OLYMPUS PK7300 AUTOMATED BLOOD BANK ANALYZER (SERIAL NUMBER (B)(4)). BOTH "REAGENT TRANSFER PITCH MOTOR ERRORS" AND "REAGENT TRANSFER FOR/BACK MOTOR ERRORS" WERE OBTAINED. THERE WAS NO PATIENT SAMPLE INVOLVED IN THIS EVENT, THEREFORE THERE WAS NO CHANGE TO OR IMPACT TO PATIENT TREATMENT IN CONJUNCTION WITH THIS EVENT. A BEC (BECKMAN COULTER) FSE (FIELD SERVICE ENGINEER) WAS DISPATCHED TO THE CUSTOMER'S LABORATORY TO ASSESS INSTRUMENT PERFORMANCE.
Patient problems: No Patient Involvement
MAUDE report 7235726 ↗
Other System, Test, Automated Blood Grouping And Antibody devices
- Galileo Echo 0087000Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody242 reports
- ORTHO 6904577Ortho-Clinical Diagnostics, Inc. · SYSTEM, TEST, AUTOMATED BLOOD GROUPING AND ANTIBODY117 reports
- Galileo Neo 0064599Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody65 reports
- TANGO infinity 850000010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody1 death55 reports
- Echo Lumena 0086998Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody49 reports
- IH-1000 001100Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody44 reports
- NEO Iris 0064597Immucor, Inc. · System, Test, Automated Blood Grouping And Antibody28 reports
- TANGO optimo 848900010Bio-Rad Medical Diagnostics GmbH · System, Test, Automated Blood Grouping And Antibody24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page