Beckman Coulter, Inc. · PK7400 Automated Microplate System · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 14987666543344), received between Jan 28, 2025 and Jan 28, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED THE PK7400 PART NUMBER (PN) B53299 SERIAL NUMBER (B)(6) HAD QUESTIONABLE FUZZY CYTOMEGALOVIRUS (CMV) RESULTS FOR SAMPLE ID (B)(6) ON (B)(6) 2025 AND THE FRONT AND BACK CONTROLS FOR CMV. THE IMAGES WERE FUZZY WITH CMV+ RESULTS. REPEAT RESULTS WERE NOT PROVIDED. SERVICE WAS NOT DISPATCHED. BECKMAN COULTER CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER OVER THE PHONE. CUSTOMER INDICATED THE ISSUE HAS BEEN RESOLVED WITH THE PROVIDED PHONE TROUBLESHOOTING. CUSTOMER INDICATED NO ISSUE WITH OTHER ASSAYS ON THE SAME PK7400. CUSTOMER REQUIRED NO FURTHER ASSISTANCE. THERE WAS NO REPORT OF DEATH OR INJURY IN CONNECTION WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21253897 ↗
  • MalfunctionJan 28, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    CUSTOMER REPORTED THE PK7400 PART NUMBER (PN) B53299 SERIAL NUMBER (B)(6) HAD QUESTIONABLE FUZZY CYTOMEGALOVIRUS (CMV) RESULT FOR SAMPLE ID (B)(6) ON (B)(6) 2024. CUSTOMER PROVIDED IMAGES OF SAMPLE ID (B)(6) AND THE FRONT AND BACK CONTROLS FOR CMV. THE IMAGES WERE FUZZY WITH CMV+ RESULTS. REPEAT RESULTS WERE NOT PROVIDED. SERVICE WAS NOT DISPATCHED. BECKMAN COULTER CUSTOMER TECHNICAL SUPPORT (CTS) ASSISTED THE CUSTOMER OVER THE PHONE. CUSTOMER INDICATED THE ISSUE HAS BEEN RESOLVED WITH THE PROVIDED PHONE TROUBLESHOOTING. CUSTOMER INDICATED NO ISSUE WITH OTHER ASSAYS ON THE SAME PK7400. CUSTOMER REQUIRED NO FURTHER ASSISTANCE. THERE WAS NO REPORT OF DEATH OR INJURY IN CONNECTION WITH THIS EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21253878 ↗

Other System, Test, Automated Blood Grouping And Antibody devices

All System, Test, Automated Blood Grouping And Antibody devices →