Portex 100/179/080Recalls & reported problems

Icu Medical, Inc. · Class 1

All models of Portex (408)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15019315022142), received between Mar 26, 2018 and Dec 23, 2025 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak5
  • Deflation Problem2
  • Leak/Splash2

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 23, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT APPROXIMATELY TWO HOURS AFTER INTUBATION, LEAKAGE WAS OBSERVED FROM THE CUFF, AND THE DEVICE WAS SUBSEQUENTLY REPLACED. THE EVENT INVOLVED A PATIENT, BUT NO INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23878340 ↗
  • MalfunctionDec 3, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT LEAKAGE FROM THE CUFF WAS OBSERVED SEVERAL TIMES APPROXIMATELY 2 HOURS AFTER INTUBATION. THE TUBE IN THE PATIENT WAS CHANGED AS A RESULT OF THE LEAKAGE. THE SAME HEALTH CARE PROFESSIONALS DID ALL THE PROCEDURES, AND THE INFLATION OF THE CUFF WAS DONE AS ALWAYS. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23704924 ↗
  • MalfunctionDec 3, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED THAT LEAKAGE FROM THE CUFF WAS OBSERVED SEVERAL TIMES APPROXIMATELY 2 HOURS AFTER INTUBATION. THE TUBE IN THE PATIENT WAS CHANGED AS A RESULT OF THE LEAKAGE. THE SAME HEALTH CARE PROFESSIONALS DID ALL THE PROCEDURES, AND THE INFLATION OF THE CUFF WAS DONE AS ALWAYS. THERE WAS PATIENT INVOLVEMENT, AND NO PATIENT HARM/ADVERSE EVENT OCCURRED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23704892 ↗
  • MalfunctionNov 24, 2025Deflation Problem; Gas/Air Leak; Deflation Problem; Gas/Air Leak

    IT WAS REPORTED THAT THE DEVICE'S PROBE AND THE BALLOON GAUGE WERE CHECKED FOR RE-INFLATION. IN THE MORNING, THERE WAS DOUBT ABOUT THE DEFLATION, SEVERAL MANOMETERS WERE USED FOR TESTING, ALL SHOWING THE SAME RESULT OF DEFLATION. DECISION OF THE ANESTHETIST DOCTOR WAS TO RE-INTUBATE THE PATIENT. DURING VISUAL INSPECTION, SMALL AIR BUBBLES WERE SEEN. AFTER CLEANING OF THE DEVICE IN A BATH AND INFLATING THE BALLOON, AIR BUBBLES WERE AGAIN OBSERVED AT THE JUNCTION BETWEEN THE TUBE AND THE BALLOON, CONFIRMING A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23633987 ↗
  • MalfunctionNov 24, 2025Deflation Problem; Gas/Air Leak; Deflation Problem; Gas/Air Leak

    IT WAS REPORTED THAT THE DEVICE'S PROBE AND THE BALLOON GAUGE WERE CHECKED FOR RE-INFLATION. IN THE MORNING, THERE WAS DOUBT ABOUT THE DEFLATION, SEVERAL MANOMETERS WERE USED FOR TESTING, ALL SHOWING THE SAME RESULT OF DEFLATION. DECISION OF THE ANESTHETIST DOCTOR WAS TO RE-INTUBATE THE PATIENT. DURING VISUAL INSPECTION, SMALL AIR BUBBLES WERE SEEN. AFTER CLEANING OF THE DEVICE IN A BATH AND INFLATING THE BALLOON, AIR BUBBLES WERE AGAIN OBSERVED AT THE JUNCTION BETWEEN THE TUBE AND THE BALLOON, CONFIRMING A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23633839 ↗
  • InjuryMar 26, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE SMITHS MEDICAL DEVICE HAD OBSERVED LEAKING BETWEEN THE TUBE AND THE CUFF. THIS ISSUE WAS OBSERVED AFTER 45 MINUTES OF USE. A TUBE CHANGE WAS REQUIRED TO RESOLVE THE ISSUE. NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7371440 ↗
  • MalfunctionMar 26, 2018Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE SMITHS MEDICAL DEVICE HAD OBSERVED LEAKING BETWEEN THE TUBE AND THE CUFF. THIS ISSUE WAS OBSERVED AFTER 24 HOURS OF USE. A TUBE CHANGE WAS REQUIRED TO RESOLVE THE ISSUE. NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7371437 ↗

FDA clearance and classification

The FDA regulates product code NTR (immunohistochemical reagent, antibody (monoclonal or polyclonal) to p63 protein in nucleus of prostatic basal cells) under 21 CFR 864.1860 as a Class I device. 21 CFR 864.1860 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Portex 100/179/080

Has the Portex 100/179/080 been recalled?

No FDA recall record in our data names this exact device identifier (DI 15019315022142), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Portex 100/179/080?

The FDA's MAUDE database lists 7 adverse event reports for this device, received between Mar 26, 2018 and Dec 23, 2025. In those reports, the most common specific device problems are “Gas/Air Leak” (5), “Deflation Problem” (2) and “Leak/Splash” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Portex 100/179/080?

None of the reports are classified as deaths, 1 as an injury and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Portex 100/179/080 FDA approved?

Devices of this type (product code NTR) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Icu Medical, Inc.

Who makes the Portex 100/179/080, and is it still sold?

The FDA's device database lists Icu Medical, Inc. as the company responsible for it (brand Portex, model 100/179/080). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMMUNOHISTOCHEMICAL REAGENT, ANTIBODY (MONOCLONAL OR POLYCLONAL) TO P63 PROTEIN IN NUCLEUS OF PROSTATIC BASAL CELLS devices

All IMMUNOHISTOCHEMICAL REAGENT, ANTIBODY (MONOCLONAL OR POLYCLONAL) TO P63 PROTEIN IN NUCLEUS OF PROSTATIC BASAL CELLS devices →