Portex 100/897/070Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Portex (409)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
1
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 15019315060434), received between Jun 5, 2017 and Aug 21, 2026 — classified as 1 death, 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Disconnection” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Disconnection2
  • Gas/Air Leak2
  • Connection Problem1
  • Detachment of Device or Device Component1
  • Device Dislodged or Dislocated1
  • Fluid/Blood Leak1
  • Improper or Incorrect Procedure or Method1
  • Inadequacy of Device Shape and/or Size1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 21, 2026Gas/Air Leak

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50379893 ↗
  • MalfunctionJul 3, 2026Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE LOCKING MECHANISM ON THE TRACHEOSTOMY TUBE DID NOT HOLD THE TUBE SECURELY IN PLACE. ON SEVERAL OCCASIONS, THE TUBE SLIPPED OUTWARD, REQUIRING READJUSTMENT AND REPOSITIONING. THIS TRACHEOSTOMY TUBE WAS INSERTED ON (B)(6) 2026. ON (B)(6) 2026, IT WAS NOTED THAT THE TRACHEOSTOMY TUBE HAD SLIPPED OUT AND THAT A GREATER PORTION OF THE TUBE WAS VISIBLE OUTSIDE THE WING. AN ENT PHYSICIAN REPOSITIONED THE TUBE BY ADVANCING IT 2¿3 CM. IT WAS OBSERVED THAT THE TUBE WAS NOT FULLY SECURED BUT APPEARED TO BE HELD MORE FIRMLY WHEN THE BLUE LOCKING MECHANISM WAS PRESSED COMPLETELY TO THE BOTTOM RATHER THAN ONLY TO THE SAME LEVEL AS THE WHITE PLASTIC COMPONENT. ON (B)(6) 2026, IT WAS AGAIN OBSERVED THAT THE TUBE HAD MIGRATED RELATIVE TO THE WING. THE TUBE WAS REPOSITIONED DOWNWARD. ON (B)(6) 2026, THE TUBE HAD ONCE AGAIN MOVED UPWARD RELATIVE TO THE WING BY APPROXIMATELY 2¿3 CM, AND IT BECAME EVIDENT THAT THE LOCKING MECHANISM WAS NOT FUNCTIONING PROPERLY. THE DEVICE WAS REPLACED WITH A NEW PORTEX UNIPERC TRACHEOSTOMY TUBE FROM A DIFFERENT LOT NUMBER. THERE WAS PATIENT INVOLVEMENT BUT NO INJURY. HOWEVER, SEVERAL DISTRESSING EPISODES OCCURRED DURING WHICH BREATHING FUNCTION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25703610 ↗
  • MalfunctionMar 25, 2026Gas/Air Leak

    IT WAS REPORTED THAT UPON INSERTION A LEAK WAS DISCOVERED. THE LEAK WAS FOUND AT THE TOP OF THE TRACHEOSTOMY TUBE AFTER REMOVAL. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688632 ↗
  • MalfunctionFeb 2, 2026Fluid/Blood Leak; Inflation Problem

    IT WAS REPORTED THAT THE CUFF WAS LEAKING AND COULD NOT INFLATE. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24239091 ↗
  • DeathJul 21, 2025Improper or Incorrect Procedure or Method; Installation-Related Problem; Improper or Incorrect Procedure or Method; Installation-Related Problem

    IT WAS REPORTED THAT THE DEVICE BECAME DISPLACED FOLLOWING SURGICAL INSERTION IN ONE OF THE ICU PATIENTS. THE MECHANISM/'CLIP' USED TO SECURE THE TRACHEOSTOMY TUBE ITSELF TO THE FLANGE CAUSED A POTENTIAL SAFETY CONCERN. THE TRACHEOSTOMY WAS MISTAKENLY INSERTED UPSIDE DOWN, WITH THE "CLIP" FACING UPWARDS. THIS RESULTED IN IT BECOMING KNOCKED OPEN WHEN THE PATIENT WAS MOVED. THE TRACHEOSTOMY BECAME DISPLACED AND THE PATIENT RAPIDLY DESATURATED DIFFICULTIES REPLACING IT THROUGH A FRESH STOMA. THE TRACHEOSTOMY BECAME DISPLACED THE MORNING AFTER IT WAS SURGICALLY PLACED ON (B)(6) 2025. IT WAS ALSO DIFFICULT TO RE-INTUBATE HER, BUT SURGICAL ASSISTANCE WAS ON HAND TO PLACE A SMALL (SIZE 6) ENDOTRACHEAL TUBE THROUGH THE TRACHEOSTOMY STOMA PRIOR TO IT BEING FORMALLY REPLACED IN THEATRE. THE PATIENT SUBSEQUENTLY DIED BUT HER DEATH WAS NOT DIRECTLY RELATED TO THE DISPLACED TRACHEOSTOMY. THE STATUS OF THE EVENT IS RESOLVED. THERE WAS PATIENT INVOLVEMENT AND ADVERSE HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22554746 ↗
  • MalfunctionNov 1, 2023Disconnection; Loose or Intermittent Connection; Connection Problem; Disconnection; Loose or Intermittent Connection; Connection Problem

    IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE IS NOT BEING HELD SECURELY BY THE FLANGE. DESPITE BEING LOCKED, THE TUBE SLIPPED THROUGH AND ADVANCED 3.5CM INTO THE PATIENT'S AIRWAY. ALSO NOTED THAT THE CONNECTOR AT THE END OF THE TRACHEOSTOMY TUBE (TO CONNECT TO CIRCUIT) IS LOOSE AND DISCONNECTS EASILY. NO INJURY REPORTED. PATIENT INVOLVEMENT. PATIENT INFORMATION IS NOT AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18047885 ↗
  • InjuryOct 19, 2023Disconnection; Detachment of Device or Device Component; Disconnection; Detachment of Device or Device Component

    IT WAS REPORTED THAT ON (B)(6) 2023, A TRACHEOSTOMY TUBE WAS PLACED IN THE PATIENT DURING AN OR PROCEDURE. ON (B)(6) 2023, THE BEDSIDE RRT NOTICED THAT THE WHITE ADAPTER OF THE TRACH TUBE WAS LOOSE AND DETACHING FROM THE TRACH TUBE (CLEAR PLASTIC TRACH TUBE SLID OFF OF TRACH ADAPTOR). MULTIPLE ATTEMPTS WERE MADE TO REATTACH THE ADAPTER BUT IT WOULD NOT STAY IN PLACE. BECAUSE THE PATIENT REQUIRED MECHANICAL VENTILATION, THE RRT HAD TO STAY AT THE BEDSIDE TO HOLD THE TRACH AND ADAPTER TOGETHER UNTIL HELP ARRIVED TO CHANGE OUT THE TUBE. THE INTENSIVIST WAS PROMPTLY NOTIFIED OF THE SITUATION AND ARRANGEMENT WERE MADE TO EXCHANGE THE TUBE UNDER VISUALIZATION WITH A FIBEROPTIC BRONCHOSCOPE. THE DEFECTIVE TRACH WAS REMOVED OVER THE BRONCH TO ENSURE AIRWAY PATENCY AND INTEGRITY. ONCE THIS WAS ESTABLISHED, A NEW TRACHEOSTOMY TUBE FROM A DIFFERENT MANUFACTURER WAS INSERTED WITHOUT INCIDENT AND PLACEMENT WAS ASSESSED AND VERIFIED WITH BRONCHOSCOPE. "NO SERIOUS HARM" REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17966132 ↗
  • MalfunctionNov 18, 2019Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size

    INFORMATION WAS RECEIVED THAT AFTER EIGHT HOURS IN USE, A SMITHS MEDICAL UNIPERC ADJUSTABLE FLANGE TRACHEOSTOMY HAD AN INSUFFICIENT SEAL. AS A RESULT, THE TUBE WAS WITHDRAWN AND A NEW TUBE WAS INSERTED. A BULGE WAS NOTED ON THE TUBE THAT WAS WITHDRAWN. NO ADVERSE EFFECTS OCCURRED,

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9336970 ↗
  • MalfunctionJun 5, 2017Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THERE WAS A LEAK IN THE "PIPE" CONNECTING THE SYRINGE TO THE BALLOON OF A PORTEX® UNIPERC® ADJUSTABLE FLANGE EXTENDED-LENGTH TRACHEOSTOMY TUBE. NO INJURY WAS REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6612183 ↗

Other Tube, tracheostomy (w/wo connector) devices