Portex 101/870/090CZRecalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 15019517076240), received between Feb 27, 2021 and Dec 20, 2023 — classified as 5 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Gas/Air Leak” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Gas/Air Leak9
- Defective Component4
- Component Incompatible1
- Disconnection1
- Inflation Problem1
- Leak/Splash1
Patient problems
- Aspiration Pneumonitis5
- Insufficient Information5
- No Clinical Signs, Symptoms or Conditions5
- No Information1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryDec 20, 2023Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
IT WAS REPORTED THAT THE DEVICE WAS MALFUNCTIONING DURING USE. THE ISSUES INCLUDE THE MALFUNCTIONING OF THE SUBGLOTTIC SUCTION VALVE ON THE NEWLY INSERTED TRACHEOSTOMY TUBE. ADDITIONALLY, THE CUFF OF THE TRACHEOSTOMY TUBE EXHIBITED IMMEDIATE LEAKAGE AFTER INSERTION. THIS PROMPTED THE NEED FOR MULTIPLE TUBE CHANGES. THIS SITUATION IS OF PARTICULAR CONCERN AS THE PATIENT UNDERGOING THESE CHANGES IS AT HIGH RISK OF ASPIRATION. IT WAS REPORTED THAT THE FOLLOWING MEASURES WERE TAKEN IN ORDER TO "SOLVE THE PROBLEM: TK CHANGE, SURVEY OF LABORATORY PARAMETERS AND ANTIBIOSIS." THERE WAS PATIENT INVOLVEMENT. THE SAMPLES ARE NOT AVAILABLE FOR EVALUATION. THE PRODUCTS WERE FOUND TO BE DEFECTIVE BETWEEN 1-NOV-2023 AND 14-NOV-2023. ONLY ONE PATIENT WAS INVOLVED. (PATIENT IDENTIFIER (B)(6).
Patient problems: Aspiration Pneumonitis; Insufficient Information
MAUDE report 18365602 ↗InjuryDec 20, 2023Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
IT WAS REPORTED THAT THE DEVICE WAS MALFUNCTIONING DURING USE. THE ISSUES INCLUDE THE MALFUNCTIONING OF THE SUBGLOTTIC SUCTION VALVE ON THE NEWLY INSERTED TRACHEOSTOMY TUBE. ADDITIONALLY, THE CUFF OF THE TRACHEOSTOMY TUBE EXHIBITED IMMEDIATE LEAKAGE AFTER INSERTION. THIS PROMPTED THE NEED FOR MULTIPLE TUBE CHANGES. THIS SITUATION IS OF PARTICULAR CONCERN AS THE PATIENT UNDERGOING THESE CHANGES IS AT HIGH RISK OF ASPIRATION. IT WAS REPORTED THAT THE FOLLOWING MEASURES WERE TAKEN IN ORDER TO "SOLVE THE PROBLEM: TK CHANGE, SURVEY OF LABORATORY PARAMETERS AND ANTIBIOSIS." THERE WAS PATIENT INVOLVEMENT. THE SAMPLES ARE NOT AVAILABLE FOR EVALUATION. THE PRODUCTS WERE FOUND TO BE DEFECTIVE BETWEEN 1-NOV-2023 AND 14-NOV-2023. ONLY ONE PATIENT WAS INVOLVED. (PATIENT IDENTIFIER (B)(6).
Patient problems: Aspiration Pneumonitis; Insufficient Information
MAUDE report 18365599 ↗InjuryDec 20, 2023Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
IT WAS REPORTED THAT THE DEVICE WAS MALFUNCTIONING DURING USE. THE ISSUES INCLUDE THE MALFUNCTIONING OF THE SUBGLOTTIC SUCTION VALVE ON THE NEWLY INSERTED TRACHEOSTOMY TUBE. ADDITIONALLY, THE CUFF OF THE TRACHEOSTOMY TUBE EXHIBITED IMMEDIATE LEAKAGE AFTER INSERTION. THIS PROMPTED THE NEED FOR MULTIPLE TUBE CHANGES. THIS SITUATION IS OF PARTICULAR CONCERN AS THE PATIENT UNDERGOING THESE CHANGES IS AT HIGH RISK OF ASPIRATION. IT WAS REPORTED THAT THE FOLLOWING MEASURES WERE TAKEN IN ORDER TO "SOLVE THE PROBLEM: TK CHANGE, SURVEY OF LABORATORY PARAMETERS AND ANTIBIOSIS." THERE WAS PATIENT INVOLVEMENT. THE SAMPLES ARE NOT AVAILABLE FOR EVALUATION. THE PRODUCTS WERE FOUND TO BE DEFECTIVE BETWEEN 1-NOV-2023 AND 14-NOV-2023. ONLY ONE PATIENT WAS INVOLVED. (PATIENT IDENTIFIER (B)(6).
Patient problems: Aspiration Pneumonitis; Insufficient Information
MAUDE report 18365528 ↗InjuryDec 20, 2023Defective Component; Gas/Air Leak; Defective Component; Gas/Air Leak
IT WAS REPORTED THAT THE DEVICE WAS MALFUNCTIONING DURING USE. THE ISSUES INCLUDE THE MALFUNCTIONING OF THE SUBGLOTTIC SUCTION VALVE ON THE NEWLY INSERTED TRACHEOSTOMY TUBE. ADDITIONALLY, THE CUFF OF THE TRACHEOSTOMY TUBE EXHIBITED IMMEDIATE LEAKAGE AFTER INSERTION. THIS PROMPTED THE NEED FOR MULTIPLE TUBE CHANGES. THIS SITUATION IS OF PARTICULAR CONCERN AS THE PATIENT UNDERGOING THESE CHANGES IS AT HIGH RISK OF ASPIRATION. IT WAS REPORTED THAT THE FOLLOWING MEASURES WERE TAKEN IN ORDER TO "SOLVE THE PROBLEM: TK CHANGE, SURVEY OF LABORATORY PARAMETERS AND ANTIBIOSIS." THERE WAS PATIENT INVOLVEMENT. THE SAMPLES ARE NOT AVAILABLE FOR EVALUATION. THE PRODUCTS WERE FOUND TO BE DEFECTIVE BETWEEN 1-NOV-2023 AND 14-NOV-2023. ONLY ONE PATIENT WAS INVOLVED. (PATIENT IDENTIFIER (B)(6).
Patient problems: Aspiration Pneumonitis; Insufficient Information
MAUDE report 18365508 ↗InjuryDec 13, 2023Component Incompatible; Inflation Problem; Gas/Air Leak; Component Incompatible; Inflation Problem; Gas/Air Leak
IT WAS REPORTED THAT THE HEALTH PROFESSIONAL EXPERIENCED ISSUES DURING USE WITH A PATIENT. THE ISSUES INCLUDE THE MALFUNCTIONING OF THE SUBGLOTTIC SUCTION VALVE ON THE NEWLY INSERTED TRACHEOSTOMY TUBE. ADDITIONALLY, THE CUFF OF THE TRACHEOSTOMY TUBE EXHIBITED IMMEDIATE LEAKAGE AFTER INSERTION, "PROMPTING THE NEED FOR MULTIPLE TUBE CHANGES". THIS SITUATION IS OF PARTICULAR CONCERN AS THE PATIENT UNDERGOING THESE CHANGES IS AT HIGH RISK OF ASPIRATION. AN UNUSED SAMPLE FROM THE SAME LOT IS AVAILABLE FOR INVESTIGATION. THE EVENT HAS AFFECTED THE PATIENT CARE AND A MEDICAL INTERVENTION WAS NECESSARY. DESCRIPTION OF THE MEDICAL INTERVENTION: THE CLIENT STATED ONLY "TK CHANGE, SURVEY OF LABORATORY PARAMETERS, AND ANTIBIOSIS". THE EVENT HAS OCCURRED IN HOSPITAL. IT IS UNKNOWN IF THERE WAS AN INJURY OR NOT. "4 OR 5 SAMPLES ARE AFFECTED. HOWEVER ONLY ONE SAMPLE OF THE SAME LOT NUMBER IS AVAILABLE FOR INVESTIGATION." PRODUCTS WERE AFFECTED DURING 01-NOV-2023 AND 14-NOV-2023, THEREFORE NO EXACT DATES COULD BE PROVIDED. A1: ONE PATIENT, AND FIVE PRODUCT MALFUNCTIONS WERE REPORTED. (B)(6) BOTH REPRESENT THE SAME PATIENT. THIS IS THE FIRST MALFUNCTION REPORT FOR THE RELATED COMPLAINTS.
Patient problems: Aspiration Pneumonitis; Insufficient Information
MAUDE report 18321024 ↗MalfunctionDec 22, 2022Disconnection; Gas/Air Leak; Disconnection; Gas/Air Leak
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, LEAKAGE OF AIR FROM IT WAS OBSERVED. THE CUSTOMER CHECKED THE PRODUCT AND THE PILOT BALLOON WAS FOUND DETACHED. NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16040357 ↗MalfunctionFeb 10, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, LEAKAGE OF AIR FROM THE CUFF WAS DETECTED. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13506638 ↗MalfunctionFeb 10, 2022Gas/Air Leak; Gas/Air Leak
IT WAS REPORTED THAT DURING THE USE OF THE PRODUCT, LEAKAGE OF FLUID BETWEEN THE PILOT BALLOON AND THE INFLATION LINE WAS FOUND. THE CUSTOMER THEN STOPPED USING THE PRODUCT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13506092 ↗MalfunctionJun 24, 2021Gas/Air Leak; Gas/Air Leak
INFORMATION WAS RECEIVED INDICATING THAT A SMITHS MEDICAL TRACHEOSTOMY CUFF WAS LEAKING AIR FROM THE CUFF. THERE WAS NO PATIENT INJURY AND NO REPORTED ADVERSE EVENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12059746 ↗MalfunctionFeb 27, 2021Leak/Splash; Leak/Splash
THE CUSTOMER CONFIRMED IN A PRE-USE CHECK THAT THE PRODUCT WORKED PROPERLY. HOWEVER, THREE (3) DAYS AFTER INDWELLING IT IN THE PATIENT, THE CUSTOMER NOTICED THE INFLATION LINE WAS TORN OFF, CAUSING LEAKAGE OF AIR FROM THE PRODUCT TO OCCUR. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 11390642 ↗
Other Tube, tracheostomy (w/wo connector) devices
- Portex 101/860/080CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)188 reports
- Portex 101/860/075CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)115 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Portex 101/860/070CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)60 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page