Portex 101/815/070Recalls & reported problems
Icu Medical, Inc. · Class 2
All models of Portex (409)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15019517077346), received between Mar 27, 2023 and Jan 12, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Incomplete or Missing Packaging” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Incomplete or Missing Packaging2
- Component Missing1
- Device Dislodged or Dislocated1
- Disconnection1
- Sharp Edges1
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2026Disconnection; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE TRACH WAS DISLODGED, ON PATIENT NECK WITH TRACH TIES STILL ATTACHED. NO BLEEDING WAS NOTED AROUND STOMA SITE. THE WRITER WAS ABLE TO PASS A SUCTION CATHETER THROUGH THE STOMA, SO THEY ATTEMPTED TO INSERT THE SAME SIZE 8 PORTEX TRACHEOSTOMY TUBE WITH AN OBTURATOR, BUT THE STOMA WAS TOO TIGHT, AND THE ATTEMPT WAS UNSUCCESSFUL. THE WRITER WAS UNABLE TO FIND A SIZE 7 PORTEX TRACHEOSTOMY TUBE, WHICH WAS THE SAME TYPE BUT SMALLER. THE WRITER WAS LATER ABLE TO FIND AND PLACE A SIZE 7 TRACHEOSTOMY TUBE ON THE FIRST ATTEMPT. THE PATIENT THEN BECAME TACHYPNEIC BUT WAS NOT IN DISTRESS. THE SPO2 WAS 94 PERCENT ON OPTIFLOW AT 85 PERCENT AND 50 L. THE RN PLACED RESTRAINTS ON THE PATIENT. THERE WERE PATIENT INJURY AND NO PATIENT DEATH.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 24042523 ↗MalfunctionJun 2, 2023Sharp Edges; Sharp Edges
IT WAS REPORTED THAT THE LENGTH OF THE INNER CANNULA EXCEEDED THE LENGTH OF THE TRACH LEAVING A SHARP EDGE TO COME IN CONTACT WITH THE PATIENTS TRACHEA. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17049220 ↗MalfunctionApr 4, 2023Component Missing; Component Missing
IT WAS REPORTED THAT WHEN A CUSTOMER OPENED A TRACH PACKAGING IT WAS NOTICED THAT THE PARTS OF THE INNER CONTENTS WERE MISSING. THESE WERE SEALED BOXES. THERE HAS BEEN NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16676304 ↗MalfunctionMar 27, 2023Incomplete or Missing Packaging; Incomplete or Missing Packaging
IT WAS REPORTED THAT DURING THE USE OF A TRACH, IT WAS DISCOVERED THAT THE TRACH WAS MISSING FROM THE UNOPENED OUTSIDE BOX, AND ONLY INSTRUCTIONS WERE FOUND INSIDE. THE DOCTOR SEARCHED FOR ANOTHER TRACH ON THE SHELF, BUT FOUND THE SAME SITUATION. THERE HAS BEEN NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16622298 ↗MalfunctionMar 27, 2023Incomplete or Missing Packaging; Incomplete or Missing Packaging
IT WAS REPORTED THAT DURING THE USE OF A TRACH, IT WAS DISCOVERED THAT THE TRACH WAS MISSING FROM THE UNOPENED OUTSIDE BOX, AND ONLY INSTRUCTIONS WERE FOUND INSIDE. THE DOCTOR SEARCHED FOR ANOTHER TRACH ON THE SHELF, BUT FOUND THE SAME SITUATION. THERE HAS BEEN NO REPORT OF OBSERVABLE CLINICAL SYMPTOMS OR A CHANGE IN SYMPTOMS IDENTIFIED IN THE PATIENT.
Patient problems: Insufficient Information
MAUDE report 16622244 ↗
Other Tube, tracheostomy (w/wo connector) devices
- Portex 101/860/080CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)188 reports
- Portex 101/860/075CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)115 reports
- Bivona 60SP040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall20 reports
- Bivona 60SP050Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall13 reports
- Bivona 60P040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall12 reports
- Portex 101/860/070CZIcu Medical, Inc. · Tube, tracheostomy (w/wo connector)60 reports
- Bivona 60SP045Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall8 reports
- Bivona 60SN040Icu Medical, Inc. · Tube, tracheostomy (w/wo connector)1 recall5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page