Skip to content

Portex 101/860/085Recalls & reported problems

Icu Medical, Inc. · Class 2

All models of Portex (409)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 15019517077940), received between Nov 13, 2023 and Jun 29, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 3 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak4
  • Detachment of Device or Device Component1
  • Insufficient Device Problem Information1
  • Unintended Deflation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 29, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE BLUSELECT 8.5 SUCTIONAID TRACHEOSTOMY TUBES DEVELOPED HOLES, RESULTING IN A LEAK THAT WAS OBSERVED AFTER APPROXIMATELY ONE DAY OF USE. THE DAMAGE WAS IDENTIFIED DURING REPLACEMENT OF THE TRACHEOSTOMY TUBE. THERE WERE PATIENT INVOLVEMENT AND NO ADVERSE HARM REPORTED, THE DAMAGED TRACHEOSTOMY TUBES COULD RESULT IN AN AIR LEAK, COMPROMISE VENTILATION, AND POTENTIALLY AFFECT THE PATIENT IF THE MALFUNCTION WERE TO RECUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25659118 ↗
  • MalfunctionJun 29, 2026Gas/Air Leak

    IT WAS REPORTED THAT THE BLUSELECT 8.5 SUCTIONAID TRACHEOSTOMY TUBES DEVELOPED HOLES, RESULTING IN A LEAK THAT WAS OBSERVED AFTER APPROXIMATELY ONE DAY OF USE. THE DAMAGE WAS IDENTIFIED DURING REPLACEMENT OF THE TRACHEOSTOMY TUBE. THERE WERE PATIENT INVOLVEMENT AND NO ADVERSE HARM REPORTED, THE DAMAGED TRACHEOSTOMY TUBES COULD RESULT IN AN AIR LEAK, COMPROMISE VENTILATION, AND POTENTIALLY AFFECT THE PATIENT IF THE MALFUNCTION WERE TO RECUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25659113 ↗
  • MalfunctionJun 12, 2026Gas/Air Leak

    IT WAS REPORTED THAT WHEN AIR WAS INJECTED INTO THE CUFF, IT IMMEDIATELY LEAKED OUT DURING VENTILATION AT PATIENT HOME. THERE WAS PATIENT INVOLVEMENT, WITH NO PATIENT HARM OR ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25492242 ↗
  • MalfunctionMar 11, 2026Gas/Air Leak

    THE CUSTOMER RETURNED THE DEVICE STATING, ¿THE CUFF WAS DAMAGED¿; DURING DEVICE EVALUATION A LEAK WAS FOUND CAUSED BY AN ASSEMBLY ERROR DURING MANUFACTURING. THE EVENT OCCURRED DURING THE PATIENT USE. THERE WAS NO ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24568527 ↗
  • MalfunctionFeb 5, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PILOT BALLOON HAD COME OFF. THE EVENT OCCURRED DURING THE PATIENT USE. THERE WAS NO ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24269069 ↗
  • InjuryNov 14, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE DEVICE INSERTED TO THE PATIENT, BY THE DOCTOR IN THE OPERATING ROOM WAS DEFLATED AFTER INSERTION. SUBSEQUENTLY, DEVICE WITH WHICH IT WAS EXCHANGED IN THE DEPARTMENT BY THE DOCTOR ALSO HAD A CUFF FAILURE. THE EXACT "FAILURE" WAS NOT DESCRIBED. NO PATIENT HARM. A1: ONE PATIENT, TWO PRODUCT MALFUNCTIONS. (B)(6) REPRESENT THE SAME PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18135001 ↗
  • InjuryNov 13, 2023Unintended Deflation; Unintended Deflation

    IT WAS REPORTED THAT THE DEVICE INSERTED TO THE PATIENT BY THE DOCTOR IN THE OPERATING ROOM WAS DEFLATED AFTER INSERTION. SUBSEQUENTLY, THE DEVICE WITH WHICH IT WAS EXCHANGED IN THE DEPARTMENT BY THE DOCTOR ALSO HAD A CUFF FAILURE. NO PATIENT HARM. A1: ONE PATIENT, TWO PRODUCT MALFUNCTIONS. (B)(6) BOTH REPRESENT THE SAME PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18125095 ↗

Other Tube, tracheostomy (w/wo connector) devices

All Tube, tracheostomy (w/wo connector) devices →