Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15019517116175), received between Dec 28, 2023 and Dec 28, 2023 — classified as 1 injury.

The most frequently reported specific device problem is “Overheating of Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device1

Patient problems

  • Full thickness (Third Degree) Burn1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 28, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT A PATIENT UNDERWENT A CARDIAC PROCEDURE AND SUSTAINED A 3 PERCENT BSA BURN OF HER RIGHT LEG AFTER THE LEVEL 1¿ CONVECTIVE WARMER DEVICE WAS DETACHED FROM A WARMING BLANKET. IT IS FURTHER ALLEGED THAT LATER INSPECTION FOUND THAT THE WARMING DEVICE WAS BLOWING SIGNIFICANTLY HOTTER AIR THAN SET TEMPERATURES. THE PATIENT IS REPORTED TO HAVE UNDERGONE EXCISIONAL DEBRIDEMENT AND SPLIT THICKNESS AUTOGRAFTING FOR WOUND CLOSURE TO THE BURNS ON THE RIGHT LEG.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 18411023 ↗