FS DISPOSABLE INTERFACE 590106ANRecalls & reported problems

Johnson & Johnson Surgical Vision, Inc. · 20 PROCEDURE PACK · Class 1

All models of FS DISPOSABLE INTERFACE (2)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Johnson & Johnson Surgical Vision FS disposable interface for eye surgery has 4,184 adverse event reports, with 3,833 malfunctions, 351 injuries and no deaths. The reports mostly describe weak or lost suction, plus debris or contamination on the interface. A few mention corneal scratches. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
4,174
Deaths reported
0
Injuries reported
351
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4,174 adverse event reports that list this exact device (DI 15050474534688), received between Aug 29, 2016 and Sep 30, 2026 — classified as 351 injuries, 3,823 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (2,827 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 5 reportsMay 2024: 6 reportsJun 2024: 2 reportsJul 2024: 2 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 7 reportsJan 2025: 1 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 9 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 4 reportsAug 2025: 3 reportsSep 2025: 12 reportsOct 2025: 6 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 8 reportsFeb 2026: 2 reportsMar 2026: 8 reportsApr 2026: 0 reportsMay 2026: 39 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Suction Problem2,827
  • Decrease in Suction1,057
  • Contamination117
  • Scratched Material57
  • Device Contaminated During Manufacture or Shipping49
  • Adverse Event Without Identified Device or Use Problem37
  • Appropriate Device Problem Term/Code Not Available27
  • Device Difficult to Setup or Prepare27

Patient problems

  • No Consequences Or Impact To Patient2,684
  • No Clinical Signs, Symptoms or Conditions1,138
  • No Code Available174
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available155
  • No Known Impact Or Consequence To Patient17
  • Corneal Abrasion16
  • Eye Injury7
  • Contusion6

Recent adverse event reports

The 25 most recent of 4,174 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Contamination

    The customer reported their patient interface with debris in contact with the patient, and no patient injury was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942924 ↗
  • MalfunctionSep 24, 2026Decrease in Suction

    It was reported by the customer that the patient interface (PI) had Suction loss during the raster pattern and the stromal bed showed a divet after the lift. There was no patient injury or surgical intervention reported. No further information was provided.NOTE: This report is against the Patient Interface. A separate report will be submitted for the FS Laser.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884021 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THE PATIENT WAS TREATED WITH THE IFS FEMTOSECOND LASER ON (B)(6) 2026, THE PATIENT EXPERIENCED STRIAE IN THE RIGHT EYE (OD). THE PATIENT HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE ON (B)(6) 2026. NO FURTHER INFORMATION WAS PROVIDED. PACHYMETRY OD/LEFT EYE (OS) = 524/519. PREOP BEST CORRECTED VISUAL ACUITY (BCVA) OD = 6/5. PREOP BCVA OS = 6/5. PREOP REFRACTION OD = S -1.75 C-1.0 S 170. PREOP REFRACTION OS = S-1.50 C-0.75 A 160. 1 DAY POSTOP UNCORRECTED VISUAL ACUITY (UCVA) OD = 6/5. 1 DAY POSTOP UCVA OS = 6/5. 1 POST FLAP LIFT UCVA OD = 6/5. 1 POST FLAP LIFT UCVA OS = 6/5. NOTE: THIS REPORT IS FOR THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25332492 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THE PATIENT WAS TREATED WITH THE IFS FEMTOSECOND LASER ON (B)(6) 2026, THE PATIENT EXPERIENCED STRIAE IN THE RIGHT EYE (OD). THE PATIENT HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE ON THE (B)(6) 2026. NO FURTHER INFORMATION WAS PROVIDED. PACHYMETRY: OD/LEFT EYE (OS) = 548/544. PREOP BEST CORRECTED VISUAL ACUITY (BCVA) OD = 6/5. PREOP BCVA OS = 6/5. PREOP REFRACTION OD = S -6.5 C-0.75 A 155/. PREOP REFRACTION OS = S -6.5 C-0.75 A 155/S-6.25 C-0.50 A 20. 1 DAY POSTOP UNCORRECTED VISUAL ACUITY (UCVA) OD = 6/20. 1 DAY POSTOP UCVA OS = 6/7.5. 1 POST FLAP LIFT UCVA OD = 6/6. 1 POST FLAP LIFT UCVA OS = 6/5. NOTE: THIS REPORT IS AGAINST THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING FILED AGAINST THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25332247 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THE PATIENT WAS TREATED WITH THE IFS FEMTOSECOND LASER ON (B)(6) 2026, THE PATIENT EXPERIENCED STRIAE IN THE LEFT EYE (OS). THE PATIENT HAD TO RETURN TO A DIFFERENT CLINIC FOR FLAP LIFT, WASH AND RINSE ON THE (B)(6) 2026. NO FURTHER INFORMATION WAS PROVIDED. PACHYMETRY: RIGHT EYE (OD)/OS = 519/516. PREOP BEST CORRECTED VISUAL ACUITY (BCVA) OD = 6/5 . PREOP BCVA OS = 6/5. PREOP REFRACTION OD = S-1.75 C-0.25 A 40. PREOP REFRACTION OS = S-2.25 C-0.25 A 180. 1 DAY POSTOP UNCORRECTED VISUAL ACUITY (UCVA) OD = 6/6. 1 DAY POSTOP UCVA OS = 6.12. 1 POST FLAP LIFT UCVA OD = 6/12. 1 POST FLAP LIFT UCVA OS = 6/5. THIS REPORT IS FOR THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25332140 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JANUARY AND MARCH 2026, ONE PATIENT EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENT HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. NOTE: THIS REPORT IS FOR THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25326756 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JULY AND DECEMBER 2023, ONE PATIENT EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENT HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. THIS REPORT IS FOR THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25326634 ↗
  • InjuryMay 29, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2024, A PATIENT EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENT HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE THIS REPORT IS FOR THE PATIENT INTERFACE. A SEPARATE REPORT IS BEING SUBMITTED FOR THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25326599 ↗
  • InjuryMay 29, 2026Mechanical Problem

    IT WAS REPORTED THAT AFTER THE PATIENT WAS TREATED WITH THE IFS FEMTOSECOND LASER HAD TO RETURN TO THE CLINIC FOR FLAP LIFT, WASH AND RINSE. THE CLINIC REPORTED THE INTRODUCTION OF A NEW BATCH OF PATIENT INTERFACES (PI) WHICH HAD DOCKING DIFFERENCES. NO FURTHER INFORMATION WAS PROVIDED. ADDITIONAL INFORMATION HAS BEEN REPORTED WITHOUT INFORMATION AS TO WHAT PATIENT IT RELATES TO. DUE TO THIS, THE BELOW HAS BEEN ADDED TO ALL RELEVANT COMPLAINT FILES TO ENSURE EVENT CAPTURED COMPLETELY. STRIAE FLAP DISPLACEMENT. THIS REPORT IS FOR THE IFS FEMTOSECOND LASER. A SEPARATE REPORT IS BEING SUBMITTED FOR THE PATIENT INTERFACE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25326532 ↗
  • MalfunctionMay 29, 2026Decrease in Suction; Device Difficult to Setup or Prepare

    IT WAS REPORTED THAT WHILE USING THE PATIENT INTERFACE (PI) THE SURGEON FOUND THE DEVICE TO BE VERY STIFF WHICH MADE PLACEMENT CHALLENGING. DUE TO THIS, THE DEVICE WOULD NOT MAINTAIN SUCTION SO WAS REPLACED WITH A BACK UP. THE SURGEON ALSO REPORTED THE REPLACEMENT DEVICE TO BE VERY STIFF. THERE WAS NO PATIENT INJURY AND THE PROCEDURE WAS COMPLETED WITHOUT FURTHER ISSUE. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25326464 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2025, FOUR PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE FOUR PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING THREE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25325819 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2025, FOUR PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE FOUR PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING THREE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25325817 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2025, FOUR PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE FOUR PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING THREE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25325804 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) AND (B)(6) 2025, FOUR PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE FOUR PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING THREE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25325550 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25324330 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25324246 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) AND (B)(6) 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25323816 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JANUARY AND JUNE 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25323720 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) AND (B)(6) 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25323627 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN (B)(6) AND (B)(6) 2025, SIX PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN UNKNOWN EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE SIX PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING FIVE PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25323502 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JANUARY AND JUNE 2023, THREE PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE THREE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE THREE PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING TWO PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25319837 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JANUARY AND JUNE 2023, THREE PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE THREE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE THREE PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING TWO PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25319762 ↗
  • InjuryMay 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING TREATMENT WITH THE IFS FEMTOSECOND LASER BETWEEN JANUARY AND JUNE 2023, THREE PATIENTS EXPERIENCED STRIAE FLAP DISPLACEMENT IN ONE EYE. THE PATIENTS HAD TO RETURN FOR FLAP LIFT, WASH AND RINSE. NO FURTHER INFORMATION WAS PROVIDED. THROUGH FOLLOW-UP, ADDITIONAL INFORMATION WAS OBTAINED; HOWEVER, IT IS UNCLEAR TO WHICH OF THE THREE PATIENTS THIS INFORMATION PERTAINS. THEREFORE, THIS INFORMATION IS BEING INCLUDED IN ALL CORRESPONDING COMPLAINT FILES. ONE (1) PATIENT HAS LEFT EYE (OS) BEST CORRECTED VISUAL ACUITY (BCVA) PREOP 6/6 AND 6/7 POST FLAP LIFT. THE REST HAVE ACHIEVED UNCORRECTED VISUAL ACUITY (UCVA) POST FLAP LIFT 6/6 OR ABOVE. ONE WITH UCVA 6/7.5 AND BCVA OF 6/6. NO ONE LOST MORE THAN 2 LINES OF BCVA. NOTE: THIS REPORT PERTAINS TO THE PATIENT INTERFACE FOR ONE OF THE THREE PATIENTS. A SEPARATE REPORT IS BEING SUBMITTED FOR THE ASSOCIATED IFS FEMTOSECOND LASER. ADDITIONALLY, SEPARATE REPORTS ARE BEING SUBMITTED FOR THE PATIENT INTERFACE FOR THE REMAINING TWO PATIENTS, AS WELL AS THE IFS FEMTOSECOND LASER.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 25319652 ↗
  • InjuryMay 14, 2026Device Difficult to Setup or Prepare; Appropriate Device Problem Term/Code Not Available; Noise, Audible

    IT WAS REPORTED THAT WHEN USING THE PATIENT INTERFACE (PI) THERE WERE ISSUES WITH SUCTION RINGS. THEY WERE VERY STIFF, CREAKY AND DIFFICULT TO USE, WHICH AFFECTED THE MENISCUS AND EPITHELIUM (SLIP) AND CAUSED ABRASIONS IN BILATERAL EYES WHICH REQUIRED THE APPLICATION OF BCL. THE PROCEDURE WAS COMPLETED. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS 2 OF 2 REPORTED.

    Patient problems: Corneal Abrasion

    MAUDE report 25178858 ↗
  • InjuryMay 14, 2026Device Difficult to Setup or Prepare; Appropriate Device Problem Term/Code Not Available; Noise, Audible

    IT WAS REPORTED THAT WHEN USING THE PATIENT INTERFACE (PI) THERE WERE ISSUES WITH SUCTION RINGS. THEY WERE VERY STIFF, CREAKY AND DIFFICULT TO USE, WHICH AFFECTED THE MENISCUS AND EPITHELIUM (SLIP) AND CAUSED ABRASIONS IN BILATERAL EYES WHICH REQUIRED THE APPLICATION OF BCL. THE PROCEDURE WAS COMPLETED. NO FURTHER INFORMATION WAS PROVIDED. THIS REPORT IS 1 OF 2 REPORTED.

    Patient problems: Corneal Abrasion

    MAUDE report 25178831 ↗

FDA clearance and classification

The FDA regulates product code HNO (keratome, ac-powered) under 21 CFR 886.4370 as a Class I device. 21 CFR 886.4370 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the FS DISPOSABLE INTERFACE 590106AN

Has the FS DISPOSABLE INTERFACE 590106AN been recalled?

No FDA recall record in our data names this exact device identifier (DI 15050474534688), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the FS DISPOSABLE INTERFACE 590106AN?

The FDA's MAUDE database lists 4,174 adverse event reports for this device, received between Aug 29, 2016 and Sep 30, 2026. In those reports, the most common specific device problems are “Suction Problem” (2,827), “Decrease in Suction” (1,057) and “Contamination” (117); the most common patient problems are “Corneal Abrasion” (16), “Eye Injury” (7) and “Contusion” (6). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the FS DISPOSABLE INTERFACE 590106AN?

None of the reports are classified as deaths, 351 as injuries and 3,823 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the FS DISPOSABLE INTERFACE 590106AN, and is it still sold?

The FDA's device database lists Johnson & Johnson Surgical Vision, Inc. as the company responsible for it (brand FS DISPOSABLE INTERFACE, model 590106AN). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.