Skip to content

Flowtron FG100Recalls & reported problems

Arjohuntleigh AB · DVT Prevention Compression Garment - FG100 - FOOT GARMENT REGULAR · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15055982700754), received between Jan 16, 2025 and Jan 16, 2025 — classified as 2 injuries.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2

Patient problems

  • Insufficient Information2
  • Nerve Damage1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 16, 2025Use of Device Problem; Use of Device Problem

    A CUSTOMER REPORTED TO THE SWEDISH MEDICAL PRODUCTS AGENCY, THE FOLLOWING: "FG 200 HAS BEEN PUT ON SO THAT WADING IS MISSING OVER AREA FOR N. THE COURSE OF THE PERONEUS OVER THE ANKLE." "SENSORY IMPAIRMENT OVER THE BIG TOE AND ADJACENT TOE. DROP FOOT WHEN INVERSION ABILITY REDUCED." THE CUSTOMER REPORTED IN THEIR REPORT THAT THE PATIENT WAS USING GARMENT SIZE FG200. HOWEVER, LOT NUMBER PROVIDED, INDICATES THAT THE GARMENT THAT WAS USED WAS FG100. THE DIFFERENCE BETWEEN THOSE TWO GARMENTS REFER TO FOOT SIZE. FG IS SMALLER THAN FG200. INFORMATION ALONG WITH PICTOGRAMS REGARDING GARMENTS SIZE, AND HOW TO APPLY THE GARMENTS ONTO THE FOOT ARE INCLUDED IN THE PRODUCT INSTRUCTION FOR USE, GARMENT SIZING AND APPLICATION GUIDE, ALSO GARMENT HAS PRINTING SHOWING THE SIZE AND PROPER APPLICATION. THE GUIDANCE AND INSTRUCTIONS FOR USE ARE DELIVERED TO THE CUSTOMER WITH EVERY GARMENT DELIVERY. THE PHOTOGRAPHIC PIECE OF EVIDENCE PROVIDED SHOWS THAT THE GARMENT WAS PLACED ON THE HEEL PART OF THE FOOT, WHICH IS NOT IN ACCORDANCE TO THE PRODUCT LABELLING. THE FOOT GARMENT HAS FOUR FLAPS, TWO FLAPS ARE FOLDED OVER THE SINGLE FLAP ON THE TOP OF THE FOOT. THE FOURTH FLAT GOES AROUND THE HEEL. THE CUSTO

    Patient problems: Insufficient Information

    MAUDE report 21172527 ↗
  • InjuryJan 16, 2025Use of Device Problem; Use of Device Problem

    A CUSTOMER REPORTED TO THE SWEDISH MEDICAL PRODUCTS AGENCY, THE FOLLOWING: "FG 200 HAS BEEN PUT ON SO THAT WADING IS MISSING OVER AREA FOR N. THE COURSE OF THE PERONEUS OVER THE ANKLE." "SENSORY IMPAIRMENT OVER THE BIG TOE AND ADJACENT TOE. DROP FOOT WHEN INVERSION ABILITY REDUCED." THE CUSTOMER REPORTED IN THEIR REPORT THAT THE PATIENT WAS USING GARMENT SIZE FG200. HOWEVER, LOT NUMBER PROVIDED, INDICATES THAT THE GARMENT THAT WAS USED WAS FG100. THE DIFFERENCE BETWEEN THOSE TWO GARMENTS REFER TO FOOT SIZE. FG IS SMALLER THAN FG200. INFORMATION ALONG WITH PICTOGRAMS REGARDING GARMENTS SIZE, AND HOW TO APPLY THE GARMENTS ONTO THE FOOT ARE INCLUDED IN THE PRODUCT INSTRUCTION FOR USE, GARMENT SIZING AND APPLICATION GUIDE, ALSO GARMENT HAS PRINTING SHOWING THE SIZE AND PROPER APPLICATION. THE GUIDANCE AND INSTRUCTIONS FOR USE ARE DELIVERED TO THE CUSTOMER WITH EVERY GARMENT DELIVERY. THE PHOTOGRAPHIC PIECE OF EVIDENCE PROVIDED SHOWS THAT THE GARMENT WAS PLACED ON THE HEEL PART OF THE FOOT, WHICH IS NOT IN ACCORDANCE TO THE PRODUCT LABELLING. THE FOOT GARMENT HAS FOUR FLAPS, TWO FLAPS ARE FOLDED OVER THE SINGLE FLAP ON THE TOP OF THE FOOT. THE FOURTH FLAT GOES AROUND THE HEEL. THE CUSTO

    Patient problems: Insufficient Information; Nerve Damage

    MAUDE report 21166610 ↗

FDA clearance and classification

Cleared through 510(k) K965153 on Jan 26, 1998 (decision: Substantially Equivalent). Submitted by Huntleigh Healthcare, Inc. as “Huntleigh Fp5000 System Model Fp5000”. The FDA review took 399 days. FDA record ↗

The FDA regulates product code JOW (sleeve, limb, compressible) under 21 CFR 870.5800 as a Class II device. 21 CFR 870.5800 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Flowtron FG100

Has the Flowtron FG100 been recalled?

No FDA recall record in our data names this exact device identifier (DI 15055982700754), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Flowtron FG100?

The FDA's MAUDE database lists 2 adverse event reports for this device, received on Jan 16, 2025. In those reports, the most common specific device problem is “Use of Device Problem” (2); the most common patient problem is “Nerve Damage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Flowtron FG100?

None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Flowtron FG100 FDA approved?

It was cleared through 510(k) K965153 on Jan 26, 1998. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Flowtron FG100, and is it still sold?

The FDA's device database lists Arjohuntleigh AB as the company responsible for it (brand Flowtron, model FG100). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SLEEVE, LIMB, COMPRESSIBLE devices

All SLEEVE, LIMB, COMPRESSIBLE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page