Skip to content

AU OSR61205Recalls & reported problems

Beckman Coulter, Inc. · UIBC · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K041793); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Jun 26, 2025Open, Classified

    Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

    Z-2264-2025 · Root cause: Process design · FDA record ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K041793. One submission often covers many models, so check the product and lot details in each record.

  • Z-3028-2020 · Aug 18, 2020 · Terminated
    UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsatu

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page