Access GI Monitor 387687Recalls & reported problems
Beckman Coulter, Inc. · The Access GI Monitor assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of CA 19-9 antigen levels in human serum and plasma using the Access Immunoassay Systems. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15099590231842), received between Jul 19, 2022 and Feb 10, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “High Test Results” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K033038); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- High Test Results2
Patient problems
- Insufficient Information2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryFeb 10, 2023High Test Results; High Test Results
THE CUSTOMER REPORTED THE PATIENT WAS ORDERED IMAGING TESTS BECAUSE OF THE HIGH ACCESS GI MONITOR RESULT. ON (B)(6)2023, THE CUSTOMER DID INDICATE THAT THE IMAGING TESTS WERE PET (POSITRON EMISSION TOMOGRAPHY) SCAN (FDG 1 -F) AND TOMOGRAPHY WITHOUT CONTRAST THAT HAVE DEMONSTRATED A DECREASE IN LESION. NO ADDITIONAL IMPACT TO THE PATIENT WAS REPORTED IN CONNECTION WITH THE EVENT.
Patient problems: Insufficient Information
MAUDE report 16348868 ↗InjuryJul 19, 2022High Test Results; High Test Results
ON (B)(6) 2022 THE CUSTOMER REPORTED OBTAINING NON-REPRODUCIBLE FALSE ELEVATED ACCESS GI MONITOR (CA 19-9) (LOT NUMBER 125013) RESULTS INVOLVING THE LABORATORY'S UNICEL DXI 800 ACCESS IMMUNO ANALYZER (SERIAL NUMBER (B)(4). THE EXACT NUMBER OF PATIENTS INVOLVED IN THIS EVENT WAS NOT SPECIFIED. THE CUSTOMER REPORTED THAT ON (B)(6) 2022, MULTIPLE PATIENTS SAMPLES RESULTS WERE AROUND 200 U/ML (AROUND LAST TWENTY TESTS OF THE REAGENT PACK 8600). THE ELEVATED PATIENT SAMPLES WERE RECHECKED WITH LOWER RESULTS. (NO DATA PROVIDED). THE CUSTOMER ALSO RECHECKED OTHER SAMPLES TESTED WITH DIFFERENT REAGENT PACKS BEFORE AND AFTER THE QUESTIONED ONE. THE RESULTS WERE AS EXPECTED. THE EXPECTED VALUES FOR THE GI MONITOR (CA 19.9) ASSAY ARE < 35 U/ML. THE CUSTOMER REPORTED SOME PATIENTS WERE ORDERED A CONTRAST ENHANCED COMPUTED TOMOGRAPHY BECAUSE OF THE HIGH ACCESS GI MONITOR RESULTS. NO ADDITIONAL IMPACT OR CHANGE TO PATIENT TREATMENT WAS REPORTED IN CONNECTION WITH THE EVENT. NO HARDWARE ERRORS OR ISSUES WITH OTHER ASSAYS WERE REPORTED IN CONJUNCTION WITH THIS EVENT. SYSTEM CHECK PASSED ON 6JUL2022. CALIBRATION PASSED ON 7JUL2022 USING REAGENT LOT 125013 AND CALIBRATOR LOT 124563. QUALITY CONTROL
Patient problems: Insufficient Information
MAUDE report 15049658 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K033038. One submission often covers many models, so check the product and lot details in each record.
- Z-2241-2020 · Apr 30, 2019 · TerminatedBeckman Coulter ACCESS Immunoassay Systems, Access GI Monitor Cancer Antigen 19-9, REF 387687 ( UDI: 15099590231842) the Access Family of Immunoassay Systems including: Access 2,
- Z-2638-2017 · May 9, 2017 · TerminatedAccess System-GI Monitor (CA19-9), A83873D, Catalog No. 387687
Other System, Test, Carbohydrate Antigen (Ca19-9), For Monitoring And Management Of Pancreatic Cancer devices
- ADVIA Centaur® CA 19-9™ 10491244Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall87 reports
- Atellica IM CA 19-9 10995489Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall73 reports
- Atellica IM CA 19-9 10995490Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall48 reports
- ADVIA Centaur® CA 19-9™ 10491379Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall37 reports
- Atellica IM CA 19-9 11206239Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER1 recall0 reports
- Elecsys CA 199 Immunoassay 11776193122Roche Diagnostics GmbH · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER32 reports
- Elecsys CA 19-9 07027028190Roche Diagnostics GmbH · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER32 reports
- Dimension Vista® Flex® reagent cartridge CA19-9 10464358Siemens Healthcare Diagnostics Inc. · SYSTEM, TEST, CARBOHYDRATE ANTIGEN (CA19-9), FOR MONITORING AND MANAGEMENT OF PANCREATIC CANCER6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page