ClearLLab 10C B96805Recalls & reported problems

Beckman Coulter Ireland Inc. · ClearLLab 10C B Cell Tube · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 15099590701475), received between Mar 25, 2022 and Oct 5, 2023 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 5, 2023Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    THE CUSTOMER REPORTED UNEXPECTED CD19+ EVENTS INTERMITTENTLY WHEN RUNNING CLEARLLAB ON THE NAVIOS EX 10 COLORS/3 LASER FLOW CYTOMETER. THE CUSTOMER REPORTED THAT THE RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO REPORT OF DEATH OR IMPACT TO PATIENT TREATMENT OR HEALTH. PATIENT RESULTS WITH QUESTIONABLE AND OR ABNORMAL SCATTER PATTERNS WERE RERUN WITH AN ALTERNATE METHOD. THE ISSUE IS NOT OCCURRING ON ALL SAMPLES. THE OTHER APPLICATIONS ARE RUNNING WITHOUT ISSUE ON THE SAME FLOW CYTOMETER. THE CUSTOMER USES THEIR OWN LABORATORY DEVELOPED SAMPLE PREPARATION PROCEDURE INSTEAD OF THE PROCEDURE IN THE INSTRUCTIONS FOR USE (IFU PN C00197). THE CLEARLAB TUBES WERE RUN ON THE NAVIOS EX 10 COLORS/3 LASER FLOW CYTOMETER: PART NUMBER: BB0910. SERIAL NUMBER: (B)(6). INSTRUMENT UDI: (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17877275 ↗
  • MalfunctionMar 25, 2022Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    WE ARE USING CLEARLLAB 10C FOR CHARACTERIZATION B CELLS LEUKEMIA. CD38 MARKER IS DETECTED IN THE CHANNEL APC-AF700. WHEN WE SETUP THE COMPENSATION WITH THE PARAMETERS DEFINED BY MANUFACTUR THE SIGNAL IN CD38 APC AF700 IS SO HIGH AND THE EVENTS ARE SHOWN IN THE EXTREM E RIGHT AXIS. WE DETECT A SPECTRAL SPILLOVER BECAUSE THE PLASMA CELLS ARE POSITIVE IN THE NEXT CHANNEL WITH THE MARKER CD20 APC AF750. ERRONEOUS PATIENT RESULTS WERE GENERATED BUT WERE NOT REPORTED OUT OF THE LABORATORY. THERE WAS NO CHANGE OR EFFECT TO PATIENT TREATMENT IN CONNECTION TO THE EVENT. THE CUSTOMER CONSIDERED THE RESULTS TO BE ERRONEOUS AND REPEATED THE TESTS USING LIQUID REAGENT TO OBTAIN ACCEPTABLE RESULTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13902717 ↗

Other Flow Cytometric Test System For Hematopoietic Neoplasms devices