Selenia Dimensions ASY-07460Recalls & reported problems
Hologic, Inc. · The Selenia® Dimensions® digital mammography system is a mammographic system used for capturing digital x-ray images of the breast. · Class 2
All models of Selenia Dimensions (9)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 15420045505605), received between Oct 12, 2017 and Jan 16, 2018 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Poor Quality Image” (3 reports).
4 other recalls cover devices cleared under the same FDA submission (P010025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Poor Quality Image3
- Device Displays Incorrect Message2
- Device Operates Differently Than Expected2
Patient problems
- Not Applicable6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 16, 2018Device Displays Incorrect Message; Device Displays Incorrect Message
IT WAS REPORTED THAT AFTER TAKING AN IMAGE A BLUE SCREEN WAS SEEN AS WELL AS AN ERROR MESSAGE. THE IMAGE WAS LOST, WHICH REQUIRED THE PATIENT TO BE RE-EXPOSED. A REBOOT WAS COMPLETED, AND AFTER THE REBOOT THE SYSTEM IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 7194756 ↗MalfunctionJan 9, 2018Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT THEY WERE GETTING GRIDLINES ON PATIENT IMAGES, CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THE KV EDGE CALIBRATION WAS OFF. GENERATOR CALIBRATIONS WERE DONE, AND THEN THE SYSTEM IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 7176230 ↗MalfunctionNov 17, 2017Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT THEY WERE GETTING GRIDLINES IN THE IMAGES CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THE MULIPLIER DRAWER NEEDED TO BE REPLACED. ONCE THIS WAS DONE, THE UNIT IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 7042604 ↗MalfunctionNov 15, 2017Poor Quality Image; Poor Quality Image
IT WAS REPORTED THAT THEY WERE GETTING GRIDLINES IN THE IMAGES CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THE GRID ASSEMBLY NEEDED TO BE REPLACED. ONCE THIS WAS DONE, THE UNIT IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 7033936 ↗MalfunctionOct 18, 2017Device Displays Incorrect Message; Device Operates Differently Than Expected; Device Displays Incorrect Message; Device Operates Differently Than Expected
IT WAS REPORTED THAT THEY WERE GETTING IMAGE CAPTURE ERRORS, CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THE GRID RETRACTION MOTOR WASN'T STRONG ENOUGH AND NEEDED REPLACED. ONCE THIS WAS DONE THE UNIT IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 6958634 ↗MalfunctionOct 12, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected
IT WAS REPORTED THAT THERE ARE LINES IN IMAGE THAT CAUSED THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THESE ARE SYMMETRICAL GATE LINES AND THE ARRAY NEEDS TO BE REPLACED. ONCE THIS WAS DONE, THE SYSTEM IS WORKING AS INTENDED.
Patient problems: Not Applicable
MAUDE report 6944745 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010025. One submission often covers many models, so check the product and lot details in each record.
- Z-2167-2026 · Apr 8, 2026 · Open, ClassifiedSELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-
- Z-1700-2014 · May 13, 2014 · TerminatedHologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography
- Z-2301-2013 · Sep 6, 2013 · TerminatedSelenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that ca
- Z-1546-2008 · Feb 28, 2008 · TerminatedSelenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
Other Full field digital,system,x-ray,mammographic devices
- Selenia Dimensions ASY-04160Hologic, Inc. · Full field digital,system,x-ray,mammographic2 recalls47 reports
- Selenia Dimensions ASY-01966Hologic, Inc. · Full field digital,system,x-ray,mammographic131 reports
- Mammomat Inspiration 10140000Siemens Healthcare GmbH · Full field digital,system,x-ray,mammographic2 recalls8 reports
- Senographe Pristina EKB3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina NAY3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB1Ge Medical Systems · Full field digital,system,x-ray,mammographic2 recalls0 reports
- Senographe Pristina NAY2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page