Selenia Dimensions ASY-08483Recalls & reported problems

Hologic, Inc. · The Selenia® Dimensions® digital mammography system generates digital mammographic images that can be used for screening and diagnosis of breast cancer. · Class 2

All models of Selenia Dimensions (9)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15420045505629), received between Sep 7, 2017 and Oct 20, 2023 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device Operates Differently Than Expected” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (P010025); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Operates Differently Than Expected2
  • Device Displays Incorrect Message1
  • Device Stops Intermittently1
  • Unintended System Motion1

Patient problems

  • Not Applicable3
  • No Clinical Signs, Symptoms or Conditions1
  • Radiation Exposure, Unintended1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 20, 2023Unintended System Motion; Unintended System Motion

    IT WAS REPORTED THAT ON (B)(6) DURING A SELENIA DIMENSIONS PROCEDURE, IT WAS REPORTED THAT THEY ATTEMPT TO PUT THE ARM AT 90° IT SOMETIMES GOES TO 45°. A FIELD ENGINEER EXAMINED THE EQUIPMENT AND DETERMINED THAT THE RIGHT ROTATION SWITCH BROKEN CAUSING UNCOMMANDED ROTATION WHEN C-ARM IS ROTATED NEGATIVELY FROM ZERO. THE EQUIPMENT WAS REPAIRED AND THE EQUIPMENT MET THE MANUFACTURER´S SPECIFICATIONS. NO INJURY TO THE PATIENT OR STAFF REPORTED. NO OTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17981120 ↗
  • MalfunctionNov 20, 2017Device Stops Intermittently; Device Stops Intermittently

    IT WAS REPORTED THAT THE SYSTEM FROZE WHEN TAKING AN IMAGE, CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED TO BE A SOFTWARE ISSUE, AND A REBOOTING THE SYSTEM RESOLVED THE ISSUE.

    Patient problems: Not Applicable

    MAUDE report 7045666 ↗
  • MalfunctionOct 9, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT THEY WERE INTERMITTENTLY GETTING LINES IN THE IMAGES CAUSING THE PATIENT TO BE RE-EXPOSED. IT WAS DETERMINED THAT THE HV POWER SUPPLY NEEDED TO BE REPLACED. ONCE THIS WAS DONE, THE UNIT IS WORKING AS INTENDED.

    Patient problems: Not Applicable

    MAUDE report 6932063 ↗
  • MalfunctionSep 27, 2017Device Operates Differently Than Expected; Device Operates Differently Than Expected

    IT WAS REPORTED THAT ONE IMAGE DID NOT PREVIEW AND WHEN THE SYSTEM WAS RESTARTED IT WOULDN'T TURN ON, THIS RESULTED IN THE ONE IMAGE BEING REPEATED. THE FE REPLACED THE COMPUTER.

    Patient problems: Not Applicable

    MAUDE report 6899827 ↗
  • MalfunctionSep 7, 2017Device Displays Incorrect Message; Device Displays Incorrect Message

    IT WAS REPORTED BY THE CUSTOMER "CAN'T CLEAR THE GEN 25:11 ERROR AND AFTER REBOOT, SAW GEN 25:8 ERROR TOO" WHICH RESULTED IN THE NEED FOR AN ADDITIONAL EXPOSURE TO COMPLETE THE EXAM. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED WHICH CONFIRMED X-RAY TUBE LARGE FILAMENT IS OPEN. THE FSE REPLACED X-RAY TUBE AND PERFORMED ALL NECESSARY CALIBRATIONS. SYSTEM PASSED ALL TESTS AND MEETS ALL MANUFACTURER'S SPECIFICATIONS.

    Patient problems: Radiation Exposure, Unintended

    MAUDE report 6849850 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P010025. One submission often covers many models, so check the product and lot details in each record.

  • Z-2167-2026 · Apr 8, 2026 · Open, Classified
    SELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-
  • Z-1700-2014 · May 13, 2014 · Terminated
    Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography
  • Z-2301-2013 · Sep 6, 2013 · Terminated
    Selenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that ca
  • Z-1546-2008 · Feb 28, 2008 · Terminated
    Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810

Other Full field digital,system,x-ray,mammographic devices

All Full field digital,system,x-ray,mammographic devices →