Selenia FFDM ASY-00676Recalls & reported problems
Hologic, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 15420045505940), received between Apr 14, 2022 and Jan 18, 2023 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Unintended System Motion” (5 reports).
4 other recalls cover devices cleared under the same FDA submission (P010025); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Unintended System Motion5
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 18, 2023Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT ON (B)(6) 2022 , DURING A SELENIA DIMENSIONS PROCEDURE THE C-ARM WAS MOVING BY ITSELF. NO PATIENT OR STAFF INJURY REPORTED. A FIELD ENGINEER EXAMINED THE EQUIPMENT AND DETERMINED THAT THE FOOT SWITCH WAS STUCK AN NEEDED TO BE REPLACED. THE SYSTEM MET THE MANUFACTURERS SPECIFICATIONS. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16195196 ↗MalfunctionJul 25, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT WHEN RELEASING THE COMPRESSION UP THE MACHINE WOULD MOVE DOWNWARD. NO INJURY REPORTED. A FIELD ENGINEER WAS DISPATCHED TO THE SITE AND DETERMINED THE FOOT PEDAL CONNECTION WAS LOOSE. THE FOOT PEDAL PINS WERE REALIGNED AND CLEANED ALLOWING FOR THE PROPER CONNECTION TO BE MADE. ONCE THIS WAS COMPLETED THE SYSTEM WAS WORKING AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15099400 ↗MalfunctionJul 18, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT THERE WAS C-ARM UNINTENDED ROTATION. NO INJURIES REPORTED. FIELD ENGINEER WAS DISPATCHED TO SITE AND AFTER INVESTIGATION THE MOTOR AND ROTOR CONTROL BOARDS WERE REPLACED. AFTER THIS THE SYSTEM WAS LEFT WORKING AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15045669 ↗MalfunctionApr 15, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT DURING A PATIENT EXAM, WHEN ATTEMPTING TO RAISE THE C-ARM, IT MOVED DOWNWARD INSTEAD AND TRAVELED DOWN UNTIL STOPPING ON ITS OWN. NO INJURY REPORTED. A FIELD ENGINEER WAS DISPATCHED TO THE SITE AND DETERMINED THE FOOT SWITCHES AND C-ARM SWITCH NEEDED TO BE REPLACED. ONCE THIS WAS COMPLETED THE SYSTEM WAS WORKING AS INTENDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14117448 ↗MalfunctionApr 14, 2022Unintended System Motion; Unintended System Motion
IT WAS REPORTED THAT THE SYSTEM C-ARM MOVED UNCONMANDED. NO INJURIES REPORTED. A FIELD ENGINEER WAS DISPATCHED TO THE SITE AND DETERMINED THAT THE DETECTOR ROTATION SWITCH NEEDED TO BE READJUSTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14112281 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P010025. One submission often covers many models, so check the product and lot details in each record.
- Z-2167-2026 · Apr 8, 2026 · Open, ClassifiedSELENIA DIMENSIONS MAMMOGRAPHY SYSTEMS. Model Numbers: SDA-SYS-3000-2D-HTC, SDA-SYS-3000-3D, SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-
- Z-1700-2014 · May 13, 2014 · TerminatedHologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography
- Z-2301-2013 · Sep 6, 2013 · TerminatedSelenia Full Field Digital Mammography System with software versions 5.0.x and 5.1.x The Selenia Full Field Digital Mammography system generates digital mammography images that ca
- Z-1546-2008 · Feb 28, 2008 · TerminatedSelenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray, mammographic; Model Number: SEL-00002; Hologic, Danbury CT 06810
Other Full field digital,system,x-ray,mammographic devices
- Selenia Dimensions ASY-04160Hologic, Inc. · Full field digital,system,x-ray,mammographic2 recalls47 reports
- Selenia Dimensions ASY-01966Hologic, Inc. · Full field digital,system,x-ray,mammographic131 reports
- Mammomat Inspiration 10140000Siemens Healthcare GmbH · Full field digital,system,x-ray,mammographic2 recalls8 reports
- Senographe Pristina EKB3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina NAY3Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB1Ge Medical Systems · Full field digital,system,x-ray,mammographic2 recalls0 reports
- Senographe Pristina NAY2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
- Senographe Pristina EKB2Ge Medical Systems · Full field digital, system, x-ray, mammographic2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.