Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 15420045508408), received between Nov 25, 2025 and Nov 25, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Contamination” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 12, 2025Open, Classified

    Fungal contamination of affected lot with Parengyodontium album.

    Z-1093-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Contamination1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 25, 2025Contamination; Contamination

    HOLOGIC SCIENTIFIC PRODUCT SUPPORT (SPS) IDENTIFIED A FUNGAL CONTAMINANT ON A SLIDE PREPARED FROM THE RETAINED QUART OF THINPREP CYTOLYT SOLUTION (QUART CONFIGURATION, LOT 5062EC) FOLLOWING A COMPLAINT FROM A CUSTOMER IN THE UNITED STATES REGARDING FUNGAL-APPEARING MATERIAL IN THE SAME LOT. DNA SEQUENCING SUBSEQUENTLY IDENTIFIED THE ORGANISM AS PARENGYODONTIUM ALBUM, AN ENVIRONMENTAL FUNGUS AND RARE OPPORTUNISTIC PATHOGEN. NO ADVERSE EVENTS, MISDIAGNOSES, OR PATIENT HARM HAVE BEEN REPORTED; HOWEVER, OUT OF AN ABUNDANCE OF CAUTION, HOLOGIC IS SUBMITTING THIS MDR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23643897 ↗

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.