InSpace™ 0131Recalls & reported problems
Ortho-Space Ltd · InSpace™ system, Medium The InSpace™ Subacromial Tissue Spacer System is a Shoulder spacer for massive irreparable rotator cuff tears, resorbable, inflatable, non-fixed. The InSpace™ Subacromial Tissue Spacer System implant is designed to act as a spacer, creating a physical barrier between tissues in the subacromial space. It is supplied sterile (EtO) and intended for single use. · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 10
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 17290013396058), received between Jun 22, 2022 and Jul 7, 2026 — classified as 10 injuries, 1 malfunction.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Patient Device Interaction Problem6
- Material Fragmentation3
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- Unspecified Infection3
- Foreign Body In Patient2
- Pain1
- Unspecified Nervous System Problem1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryJul 7, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PERSISTENT WEAKNESS OF LEFT ROTATOR CUFF, PAIN OF LEFT SHOULDER, AND ¿HARD BUMP¿ FELT IN LEFT SHOULDER AFTER APPROXIMATELY 3 MONTHS. PATIENT REQUIRED RE-ADMISSION TO HOSPITAL, MEDICAL CLEARANCE AND PROCEDURE TO REMOVE IMPLANT.
Patient problems: Pain
MAUDE report 25742100 ↗InjuryJun 23, 2026Material Fragmentation
IT WAS REPORTED THAT THE DOCTOR USED AN INSPACE BALLOON AND THE SURGERY AND PROCEDURE WENT SMOOTHLY WITH NO ISSUES. THE DOCTOR THEN INFORMED ME HE RECEIVED POST-OP X RAYS, SHOWING THE INSPACE SMALL METAL INSERTER PIECE WAS LEFT BEHIND IN THE SHOULDER JOINT.
Patient problems: Foreign Body In Patient
MAUDE report 25602123 ↗MalfunctionNov 10, 2025Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT AFTER INSERTING THE RECOMMENDED AMOUNT OF SALINE INTO A SIZE MEDIUM INSPACE BALLOON, IT WAS NOTICED THAT THE METAL TIP OF THE INSERTER HAD BROKEN OFF INTO THE BALLOON. THE BALLOON WAS THEN REMOVED AND ANOTHER BALLOON WAS INSERTED IT WAS ALSO A SIZE MEDIUM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23516135 ↗InjuryJun 13, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT 6 WEEKS AFTER THE PATIENT¿S SURGERY THE BALLOON SLID ANTERIORLY AND LATERAL. MRI WAS PERFORMED. DOCTOR PERFORMED A REVISION SURGERY ON (B)(6) AND THE BALLOON WAS COMPLETELY DEFLATED UPON SCOPING THE SHOULDER. WHEN THE BALLOON WAS REMOVED, THERE WAS A HUGE HOLE IN THE SIDE OF IT, INDICATING THAT IT HAD POPPED AT SOME POINT. THIS PROCEDURE WAS DONE ON TOP OF A REPAIR AND THE SUTURES WERE NOTICEABLY TORN WHEN WE WENT IN AS WELL. DURING THE 2ND PROCEDURE, THE DOCTOR PLACED A NEW LARGE BALLOON IN THE PATIENT WHICH WENT IN WITHOUT ISSUES. HE ALSO DID AN OPEN SUBPECTENODESIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22206834 ↗InjuryNov 6, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
DURING LAST FOLLOW UP AFTER INSPACE IMPLANTATION, THE DOCTOR DISCOVERED THE PATIENT HAS AXILLARY NERVE PALSY. THE DOCTOR IS UNSURE IF THIS IS DUE TO THE BALLOON OR ANOTHER FACTOR. THE PATIENT ALSO RECEIVED A NECK INJECTION ABOUT 6 WEEKS AFTER THE SPACER, WHICH COULD HAVE DONE DAMAGE, BUT IT¿S UNCLEAR. THE PATIENT NOW HAS PSEUDO PARALYSIS OF THE SHOULDER.
Patient problems: Unspecified Nervous System Problem
MAUDE report 20616130 ↗InjuryAug 12, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE INSPACE BALLOON IMPLANT SHIFTED POSTERIORLY IN THE SUBACROMIAL SPACE 2 WEEKS POST OP. THE BALLOON WAS DEFLATED UPON EXAMINATION PRIOR TO REMOVAL. THE BALLOON WAS REMOVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19968454 ↗InjuryJun 7, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE PATIENT HAD A SUBCUTANEOUS DEVELOPMENT OF ¿COTTAGE CHEESE¿ TYPE CONSISTENCY AFTER SURGERY.
Patient problems: Unspecified Infection
MAUDE report 19486707 ↗InjuryJun 7, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE PATIENT HAD A SUBCUTANEOUS DEVELOPMENT OF ¿COTTAGE CHEESE¿ TYPE CONSISTENCY.
Patient problems: Unspecified Infection
MAUDE report 19486602 ↗InjuryMay 3, 2024Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE INITIAL PROCEDURE HAD NO ISSUES AND WAS SUCCESSFUL AT THE TIME OF INSERTION. AT NO TIME WAS THERE ANY BELIEF THE PRODUCT WAS DAMAGED OR BROKEN AFTER INSERTION. THE PROCEDURE WAS FINISHED AND WENT WELL. ON LATER DATE, AFTER OPERATION, THE PATIENT FELL AND AGGRAVATED HIS SHOULDER. PATIENT HAD X-RAYS ON WHICH SHOWED ¿A NEW TINY MAN-MADE DENSITY PROJECTING BETWEEN THE ACROMION AND THE HUMERAL HEAD MEASURING 4 X 1 MM.¿ FURTHER X-RAYS ON THEN SHOWED ¿THERE IS A METALLIC CLIP IN THE SUBACROMIAL SPACE.¿ AN ULTRASOUND ALSO SHOWED ¿A SMALL AMOUNT OF FLUID IS IDENTIFIED IN BETWEEN THE ACROMION AND HEAD OF THE HUMERUS. MORE INFERIOR AND DEEPER, ALONG THE PROXIMAL/MEDIAL HUMERAL HEAD THERE IS A COLLECTION OF FLUID MEASURING 3.4 X 2.2 X 2.3 CM.¿ AN ADDITIONAL SURGERY WAS THEN COMPLETED BY THE SURGEON TO REMOVE THE INSPACE BALLOON WITH ¿BALLOON REMOVED INTACT WITHOUT FLUID WITH METAL FRAGMENT INSIDE IT.¿ ACCORDING TO THE SURGEON, THE TIP OF THE INFLATION CANNULA BROKE OFF IN THE NECK OF THE DEVICE. WITH THE INFLATION CANNULA NOT VISIBLE PRIOR TO THE DEPLOYMENT AND INFLATION OF THE DEVICE, STAFF DID NOT NOTICE A DIFFERENCE IN THE CANNULA.
Patient problems: Foreign Body In Patient
MAUDE report 19240164 ↗InjuryMar 22, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THERE WAS PATIENT PAIN THAT REQUIRED AN ADDITIONAL SURGERY TO REMOVE THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16591959 ↗InjuryJun 22, 2022Patient Device Interaction Problem; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION AND REQUIRED A WASH OUT.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Infection
MAUDE report 14778111 ↗
FDA clearance and classification
Granted through the De Novo pathway (DEN200039) on Jul 12, 2021. Submitted by Ortho-Space , Ltd. as “InSpace Subacromial Tissue Spacer System”. The FDA review took 395 days. FDA record ↗
The FDA regulates product code QPQ (shoulder spacer for massive irreparable rotator cuff tear, resorbable, inflatable, non-fixed) under 21 CFR 888.3630 as a Class II device. 21 CFR 888.3630 ↗
FDA definition of this device type: A resorbable shoulder spacer is intended to act as a temporary spacer, creating a physical barrier between tissues in the shoulder, for the treatment of massive irreparable rotator cuff tears.
Common questions about the InSpace™ 0131
Has the InSpace™ 0131 been recalled?
No FDA recall record in our data names this exact device identifier (DI 17290013396058), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the InSpace™ 0131?
The FDA's MAUDE database lists 11 adverse event reports for this device, received between Jun 22, 2022 and Jul 7, 2026. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (6), “Material Fragmentation” (3) and “Device Dislodged or Dislocated” (2); the most common patient problems are “Unspecified Infection” (3), “Foreign Body In Patient” (2) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the InSpace™ 0131?
None of the reports are classified as deaths, 10 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the InSpace™ 0131 FDA approved?
It was granted through the De Novo pathway (DEN200039) on Jul 12, 2021.
Is the InSpace™ 0131 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the InSpace™ 0131, and is it still sold?
The FDA's device database lists Ortho-Space Ltd as the company responsible for it (brand InSpace™, model 0131). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Shoulder Spacer For Massive Irreparable Rotator Cuff Tear, Resorbable, Inflatable, Non-Fixed devices
- InSpace™ 0132Ortho-Space Ltd · Shoulder Spacer For Massive Irreparable Rotator Cuff Tear, Resorbable, Inflatable, Non-Fixed6 reports
- InSpace™ 0130Ortho-Space Ltd · Shoulder Spacer For Massive Irreparable Rotator Cuff Tear, Resorbable, Inflatable, Non-Fixed2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page