Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1 (REF 54-11230)Recalls & reported problems
Orthofix Srl · TL+ HALF PIN AO 4MM X 180MM · Class 2 · Implant
All models of Orthofix TrueLok Hexapod System (TL-HEX) V2.0 (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 18033509858222), received between May 8, 2019 and May 8, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K170650); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 8, 2019Break
INFORMATION PROVIDED STATES THAT APPX. 2 1/2 WEEKS POST OP THE PATIENT WAS WEIGHT BEARING WITH A ROCKER RAIL ATTACHED TO THE FRAME. TWO OF THE THREE HALF PINS THAT WERE APPLIED SHEARED. THE 5MM HALF PIN BROKE 1/2 CM FROM THE FIXATION BOLT AND WAS REMOVED. THE 4MM HALF PIN BROKE AT THE SKIN AND WAS LEFT IN THE PATIENT. THE FRAME REMAINS ON THE PATIENT AND WILL BE REMOVED AFTER COMPLETION OF TREATMENT.
Patient problems: Foreign Body In Patient
MAUDE report 8593735 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K170650. One submission often covers many models, so check the product and lot details in each record.
- Z-2063-2018 · Jan 30, 2018 · TerminatedTL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of r
Other Software for diagnosis/treatment devices
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall3 reports
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall1 report
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall0 reports
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall0 reports
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall0 reports
- Orthofix TrueLok Hexapod System (TL-HEX) V2.0 1Orthofix Srl · Software For Diagnosis/Treatment1 recall0 reports
- MAXFRAME 1100007-01Synthes (U.S.A.) LP · Software for diagnosis/treatment14 reports
- MAXFRAME 1100010-01Synthes (U.S.A.) LP · Software for diagnosis/treatment10 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026