Skip to content

Quantum PureFlow Standard Heat Exchanger High Flow 3/8 W HX55V-S0WRecalls & reported problems

Spectrum Medical Srl · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 18051160302205), received between Nov 21, 2025 and May 19, 2026 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K223879); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jul 17, 2026Class IIOpen, Classified

    Due to blood-to-glycol leak in the heat exchanger.

    Z-3327-2026 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Tear, Rip or Hole in Device Packaging1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2026Fluid/Blood Leak

    BLOOD NOTED IN GLYCOL COMPARTMENT OF HEX - APPEARED TO BE A FAULT IN THE SEAM LINE BETWEEN THE GLYCOL COMPARTMENT AND THE BLOOD CHANNEL. DID NOT APPEAR TO RUPTURE, BLOOD APPEARED TO BE CONTAINED. CLINICIAN DISCARDED THE HEX FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25213370 ↗
  • InjuryDec 10, 2025Fluid/Blood Leak; Fluid/Blood Leak

    USER REPORTED A BLOOD TO GLYCOL LEAK DURING THE CASE. NO ADVERSE CONSEQUENCES OCCURRED FOR THE PATIENT.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 23772177 ↗
  • MalfunctionNov 21, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    USER REPORTED SIGNS OF DAMAGE TO THE STERILE PRIMARY PACKAGING (BLISTER) BEFORE USE. NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23616348 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K223879. One submission often covers many models, so check the product and lot details in each record.

  • Z-3328-2026 · Jul 17, 2026 · Open, Classified
    Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quan