Skip to content

CETRA 19-0236Recalls & reported problems

Orthofix US LLC · 2-LEVEL PLATE ASSEMBLY, 36MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 18257200120406), received between Dec 3, 2025 and Jul 27, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 3 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated2
  • Unintended Movement2
  • Break1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information3
  • Intervertebral Disc Compression or Protrusion1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Insufficient Device Problem Information

    It was reported a patient with new herniated nucleus pulpous (HNP) at C6/7 (below fusion) after a fall-- with cord compression.

    Patient problems: Intervertebral Disc Compression or Protrusion

    MAUDE report 50015873 ↗
  • MalfunctionDec 3, 2025Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement

    IT WAS REPORTED THAT A PATIENT RECEIVED A 36MM CETRA PLATE WITH TWO TI CAGES IN (B)(6) 2025. AT A FOLLOW UP VISIT THE PHYSICIAN NOTICED TWO INFERIOR SCREWS HAD BACKED OUT. THE PATIENT SUBSEQUENTLY UNDERWENT A REVISION OF THE ACDF ON (B)(6) 2025. DURING THE REVISION IT WAS OBSERVED THE LOCKING PLATE HAS BROKEN OFF OF THE PLATE. THE PHYSICIAN REMOVED THE ENTIRE CONSRTUCT AND REVISED COMPLETELY.

    Patient problems: Insufficient Information

    MAUDE report 23707985 ↗
  • MalfunctionDec 3, 2025Device Dislodged or Dislocated; Unintended Movement; Device Dislodged or Dislocated; Unintended Movement

    IT WAS REPORTED THAT A PATIENT RECEIVED A 36MM CETRA PLATE WITH TWO TI CAGES IN (B)(6) 2025. AT A FOLLOW UP VISIT THE PHYSICIAN NOTICED TWO INFERIOR SCREWS HAD BACKED OUT. THE PATIENT SUBSEQUENTLY UNDERWENT A REVISION OF THE ACDF (B)(6) 2025. DURING THE REVISION IT WAS OBSERVED THE LOCKING PLATE HAS BROKEN OFF OF THE PLATE. THE PHYSICIAN REMOVED THE ENTIRE CONSTRUCT AND REVISED COMPLETELY.

    Patient problems: Insufficient Information

    MAUDE report 23707896 ↗
  • MalfunctionDec 3, 2025Break; Break

    IT WAS REPORTED THAT A PATIENT RECEIVED A 36MM CETRA PLATE WITH TWO TI CAGES IN (B)(6) 2025. AT A FOLLOW UP VISIT THE PHYSICIAN NOTICED TWO INFERIOR SCREWS HAD BACKED OUT. THE PATIENT SUBSEQUENTLY UNDERWENT A REVISION OF THE ACDF ON (B)(6) 2025. DURING THE REVISION IT WAS OBSERVED THE LOCKING PLATE HAS BROKEN OFF OF THE PLATE. THE PHYSICIAN REMOVED THE ENTIRE CONSRTUCT AND REVISED COMPLETELY.

    Patient problems: Insufficient Information

    MAUDE report 23707312 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page