MozecTM NC Rx PTCA Balloon Dilatation Catheter MNC30013Recalls & reported problems
Meril Life Sciences Private Limited · MozecTM NC Rx PTCA Balloon Dilatation Catheter, 3.00 mm x 13 mm, Rapid Exchange · Class 2
All models of MozecTM NC Rx PTCA Balloon Dilatation Catheter (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 18906029358377), received between Dec 11, 2018 and Mar 9, 2020 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Deformation1
- Physical Resistance/Sticking1
- Retraction Problem1
Patient problems
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 9, 2020Material Deformation; Material Deformation
AS REPORTED, A 3.00 X 13 MM MOZECTM NC - RX PTCA BALLOON DILATATION CATHETER WAS USED FOR A POST DILATION OF A STENT. IT DID NOT POST DILATE THE DEPLOYED ONYX STENT. THE BLUE PART OF THE SHAFT LOOKS WITH DEFECT. THERE WAS NO PATIENT INJURY REPORTED. THE DEVICE WAS USED FOR THE MALE TO POST DILATE A STENT. THERE WAS NO DIFFICULTY IN REMOVING THE PRODUCT FROM THE HOOP. THERE WAS NO DIFFICULTY IN REMOVING THE PROTECTIVE BALLOON COVER OR THE STYLET. THERE WERE NO ANOMALIES NOTED TO THE DEVICE PRIOR TO USE. THE DEVICE WAS STORED AND HANDLED PER THE IFU. THE DEVICE WAS PREPPED PER THE IFU. THERE WERE NO KINKS OR OTHER DAMAGES NOTED TO THE DEVICE PRIOR TO INSERTING INTO THE PATIENT. THERE WAS NO RESISTANCE/FRICTION EXPERIENCED WHILE INSERTING THE BALLOON THROUGH THE ROTATING HAEMOSTATIC VALVE. THERE WAS NO RESISTANCE/FRICTION EXPERIENCED WHILE INSERTING THE DEVICE THROUGH THE GUIDE CATHETER. THERE WAS NO UNUSUAL FORCE USED DURING THE LOADING ON THE GUIDE WIRE. THERE WAS NO RESISTANCE OR ABNORMAL DEVICE INTERACTION NOTICED DURING ADVANCEMENT. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER THROUGH THE VESSEL. THERE WAS NO RESISTANCE OR ABNORMAL DEVICE INTERACTION NOTICED DURING LESI
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9804182 ↗MalfunctionDec 11, 2018Retraction Problem; Physical Resistance/Sticking; Retraction Problem; Physical Resistance/Sticking
IT WAS REPORTED THAT A 3.0X13 MOZEE NC WAS DIFFICULT TO REMOVE FROM SION BLUE WIRE. NO PATIENT HARM OR INJURY WAS REPORTED. THE DEVICE WAS STORED AND HANDLED AS PER THE IFU. THERE WAS NO DAMAGE NOTED TO THE PACKAGING OF THE DEVICE AND THE DEVICE WAS REMOVED FROM THE PACKAGE WITHOUT ENCOUNTERING ANY DIFFICULTY. THERE WAS NO DIFFICULTY IN REMOVING THE PRODUCT FROM THE HOOP AND IN REMOVING THE PROTECTIVE BALLOON COVER. THE PRODUCT WAS NORMALLY PREPPED FOLLOWING THE IFU WITHOUT USING ANY EXCESS FORCE. THE PRODUCT WAS INSPECTED PRIOR TO USE AND APPEARED TO BE NORMAL. NO ANOMALIES WERE NOTED DURING OR AFTER THE DEVICE WAS PREPPED. NO VISIBLE SIGNS OF KINKS OR DAMAGES WERE NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. NO RESISTANCE I FRICTION WAS EXPERIENCED WHILE INSERTING THE BALLOON THROUGH THE ROTATING HEMOSTATIC VALVE AND WHILE CROSSING THE TARGET LESION. NO VISIBLE SIGNS OF KINK OR DAMAGE WERE NOTED ANYTIME DURING THE PROCEDURE AND NO UNUSUAL FORCE WAS APPLIED ANYTIME DURING THE PROCEDURE. THERE WAS NO DIFFICULTY ENCOUNTERED WHEN CONNECTING THE HUB TO THE INDEFLATOR AND THE DEVICE WAS NOT ADVANCED WITH THE INDEFLATOR CONNECTED TO THE HUB. THE BALLOON WAS INFLATED NORMALLY A
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8150942 ↗
Other Catheters, Transluminal Coronary Angioplasty, Percutaneous devices
- MINI TREK™ 1012270-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous223 reports
- Maverick²™ H7493892815200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous1 death203 reports
- MINI TREK™ 1012270-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous162 reports
- Maverick²™ H7493892820200Boston Scientific Corporation · Catheters, transluminal coronary angioplasty, percutaneous161 reports
- NC TREK 1012454-08Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 recalls59 reports
- TREK™ 1012272-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous2 deaths158 reports
- NC TREK 1012453-12Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall84 reports
- NC TREK 1012453-15Abbott Vascular Inc. · Catheters, transluminal coronary angioplasty, percutaneous1 recall1 death83 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page