Skip to content

Clearview™ 22120Recalls & reported problems

Medtronic, Inc. · BLOWER/MISTER 22120 CLEARVIEW 10PK 17L · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20643169888672), received between Mar 27, 2024 and Aug 22, 2024 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 19, 2024Open, Classified

    Potential for unsealed sterile packing.

    Z-1794-2024 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Break1
  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2024Obstruction of Flow; Obstruction of Flow

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CLEARVIEW BLOWER MISTER, IT WAS REPORTED THAT NO CO2 GAS CAME OUT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT NONE OF THE DEVICE TUBING OR THE DISTAL TIP APPEARED TO BE DAMAGED OR OCCLUDED. THERE WAS NO DAMAGE OR KINK IN THE TUBING. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER THINKS THE TIP WAS ON THE DEVICE WHEN THEY USED IT BUT THEY ARE NOT CERTAIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20026707 ↗
  • MalfunctionMar 27, 2024Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A CLEARVIEW BLOWER/MISTER, IT WAS REPORTED THAT THE IV PART OF THE DEVICE WAS BROK EN. USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18985131 ↗

Other LAVAGE, JET devices

All LAVAGE, JET devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.