POWERFLEX 4207000SRecalls & reported problems
Cordis US Corp. · POWERFLEX P3 F5 7X10 80 · Class 2
All models of POWERFLEX (56)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032004517), received between Jul 11, 2022 and Jul 11, 2022 — classified as 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (1 report).
1 other recall cover devices cleared under the same FDA submission (K032737); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Burst Container or Vessel1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 11, 2022Burst Container or Vessel; Burst Container or Vessel
AS REPORTED, THE BALLOON OF A 7MM X 10CM 80CM POWERFLEX P3 F5 PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) BALLOON CATHETER RUPTURED BEFORE REACHING ITS NOMINAL PRESSURE. AS A RESULT, A NEW 7MM X 10CM 80CM POWERFLEX P3 BALLOON CATHETER WAS USED IN PLACE OF THE 7MM X 10CM 80CM POWERFLEX P3 PTA BALLOON CATHETER AND WAS USED WITHOUT ISSUE. THERE WAS NO REPORTED INJURY TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT A FISTULA LESION WHICH HAD AN 80% STENOSIS BUT SHOWED NO SIGNS OF CHRONIC TOTAL OCCLUSION (CTO), CALCIFICATION, OR TORTUOSITY. A NON-CORDIS .035 GUIDEWIRE WAS USED FOR THE PROCEDURE WITHOUT ISSUE. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE HOOP, REMOVING THE PROTECTIVE BALLOON COVER, THE STYLET, OR ANY OF THE STERILE PACKAGING COMPONENTS AND THE DEVICE WAS NOT KINKED OR DAMAGED IN ANY WAY PRIOR TO BEING INSERTED INTO THE PATIENT. THERE WAS NO RESISTANCE OR FRICTION DURING INSERTION OF THE DEVICE. THE DEVICE WAS ABLE TO TRACK TOWARD THE LESION AND CROSS THE LESION WITHOUT DIFFICULTY. THE DEVICE WAS NEVER PUT INTO AN ACUTE BEND AND DID NOT KI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14976318 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032737. One submission often covers many models, so check the product and lot details in each record.
- Z-1477-2020 · Jan 8, 2020 · TerminatedCordis POWERFLEX P3, PTA Dilation Catheter, REF 420-6020S
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page