CORDIS 67003500Recalls & reported problems
Cordis US Corp. · 6F .070 AL.75 SH 100CM · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032020043), received between Oct 19, 2020 and Oct 19, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Material Puncture/Hole” (1 report).
1 other recall cover devices cleared under the same FDA submission (K963962); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Puncture/Hole1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2020Material Puncture/Hole; Material Puncture/Hole
AS REPORTED, A 6F .072¿ EXTRA BACK-UP (XB) 3 100CM VISTA BRITE TIP GUIDING CATHETER WAS KINKED. ANOTHER 6F .072 100CM VISTA BRITE TIP AMPLATZ (AL).75 SIDEHOLES GUIDING CATHETER WAS KINKED WITH A SMALL HOLE IN THE KINK PART. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10697772 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K963962. One submission often covers many models, so check the product and lot details in each record.
- Z-1198-2019 · Feb 7, 2019 · TerminatedCordis Vista Brite Tip & Androit Guiding Catheters Product Usage: The guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page