VISTA BRITE TIP 77803400Recalls & reported problems
Cordis Corporation · GC 7F 078 AL .75 · Class 2
All models of VISTA BRITE TIP (26)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 20705032021699), received between Jun 23, 2020 and Jun 23, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Crack” (1 report).
1 other recall cover devices cleared under the same FDA submission (K963962); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Crack1
Patient problems
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 23, 2020Crack; Crack
AS REPORTED, THE HEAD-END OF 7F .078" .75 JAL VISTA BRITETIP GUIDING CATHETER WAS FOUND FRACTURED DURING USE. THEREFORE, THE DEVICE WAS REPLACED WITH ANOTHER UNKNOWN GUIDE CATHETER TO COMPLETE THE OPERATION. IT HAS NO EFFECT TO THE PATIENT. THERE WAS NO REPORTED PATIENT INJURY. THE PATIENT HAD HISTORY OF CORONARY HEART DISEASE AND HAD BEEN IMPLANTED WITH UNKNOWN STENT BEFORE. THE PATIENT HAD EXPERIENCED PAIN CAUSING HIS RE-HOSPITALIZATION. FOLLOWING A RADIOGRAPHIC EXAMINATION, IT WAS FOUND THAT THE LEFT CORONARY ARTERY (LCA) HAS 80% STENOSIS THAT SUGGESTED A NEED TO BE RE-TREATED WITH PERCUTANEOUS CORONARY INTERVENTION (PCI). THE DEVICE WAS STORED, HANDLED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU). THERE WAS NO VISIBLE DAMAGE NOTED TO THE DEVICE PRIOR TO OPENING THE PACKAGE. THE LESION WAS MODERATELY CALCIFIED, WITH NO VESSEL TORTUOSITY. THE DEVICE WAS NOT USED FOR A CHRONIC TOTAL OCCLUSION (CTO). THE ACCESS SITE WAS THE RADIAL ARTERY. THERE WAS NO EXCESS FORCE USED AT ANY TIME DURING THE PROCEDURE. THE HEAD-END WAS KINKED. THE DEVICE WAS REMOVED IN ONE PIECE. ANOTHER VISTA BRITE TIP GUIDING CATHETER WAS USED TO COMPLETE THE PROCEDURE. THE DEVICE WILL BE RETURNED FOR EVALUATIO
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10184335 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K963962. One submission often covers many models, so check the product and lot details in each record.
- Z-1198-2019 · Feb 7, 2019 · TerminatedCordis Vista Brite Tip & Androit Guiding Catheters Product Usage: The guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page