S.M.A.R.T. C06150MBRecalls & reported problems
Cordis US Corp. · SMART 120/150, BILIARY - 6X150 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032023693), received between Apr 19, 2018 and Oct 21, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component1
- Detachment of Device or Device Component1
- Difficult or Delayed Positioning1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE SMART 6X150 STENT WAS DEPLOYED WITHOUT ISSUES. WHEN THE STENT WAS PULLED OUT OVER THE WIRE, IT WAS NOTED THAT THERE WAS A SECOND INNER CANNULA THAT HAD SEPARATED FROM THE REST OF THE DEVICE. IT WAS RETAINED ON THE WIRE AND IN THE PATIENT. A PIECE OF THE RETAINED CANNULA WAS STICKING OUT OF THE PATIENT, IT WAS THEN REMOVED. THE PHYSICIAN NOTED THERE WAS NO HARM TO THE PATIENT. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; HOWEVER, ONE JPG. IMAGE WAS REVIEWED. A PARTIALLY IDENTIFIED SMART STENT DELIVERY SYSTEM IS CONTAINED WITHIN WHAT APPEARS TO BE THE ORIGINAL PACKAGING, WHICH HAS BEEN OPENED AND USED IN SOME WAY. THERE APPEARS TO BE TWO SEPARATE SHAFTS OR LUMENS PRESENT: THE BLUE OUTER SHEATH AND A BROWN SECONDARY SHAFT. THE STENT IS NOT APPARENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15653154 ↗InjuryApr 19, 2018Detachment Of Device Component; Difficult or Delayed Positioning; Detachment Of Device Component; Difficult or Delayed Positioning
DURING DEPLOYMENT, IT WAS REPORTED THAT THE SMART STENT WAS PARTIALLY DEPLOYED AND THE STENT DELIVERY SYSTEM SEPARATED WHEN AN ATTEMPT WAS MADE TO REMOVE IT FROM THE PATIENT. THE SDS SEPARATED IN THE PATIENT AND THE REMAINING PIECE HAD TO BE SNARED FOR REMOVAL. THERE WAS NO PATIENT INJURY. THE PHYSICIAN WAS DEPLOYING A SMART STENT AND STATED THAT THE STENT BECAME STUCK AND WOULDN'T DEPLOY CORRECTLY. A 6F UNKNOWN SHEATH AND A 0.035 UNKNOWN GUIDEWIRE WAS IN PLACE. THE PHYSICIAN HAD TO FORCEFULLY REMOVE THE STENT AND IT "BROKE OFF". THE PHYSICIAN HAD TO SNARE THE "BROKEN OFF" SECTION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7442877 ↗
Other CATHETER, PERCUTANEOUS devices
- WATCHMAN FXD Curve™ Access System M635TU80020Boston Scientific Corporation · Catheter, percutaneous1 recall47 deaths654 reports
- NA C315HIS02Medtronic, Inc. · Catheter, percutaneous1 recall369 reports
- WATCHMAN TruSteer™ Access System M635TU90050Boston Scientific Corporation · Catheter, percutaneous1 recall28 deaths302 reports
- Guidezilla™ II H7493933515060Boston Scientific Corporation · CATHETER, PERCUTANEOUS1 recall3 deaths278 reports
- Velocity Delivery Microcatheter VEL160STR-APenumbra, Inc. · CATHETER, PERCUTANEOUS297 reports
- Guidezilla™ H7493924215050Boston Scientific Corporation · CATHETER, PERCUTANEOUS294 reports
- WATCHMAN™ TruSeal™ Access System M635TU70020Boston Scientific Corporation · Catheter, percutaneous1 recall11 deaths217 reports
- Lantern Delivery Catheter PXSLIMLAN135T45Penumbra, Inc. · CATHETER, PERCUTANEOUS220 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page