Skip to content

Cordis PRECISE PRO Rx Nitinol Stent System PC0730RXCRecalls & reported problems

Cordis US Corp. · Class 3 · Implant

All models of Cordis PRECISE PRO Rx Nitinol Stent System (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032036457), received between Mar 30, 2017 and Nov 24, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

4 other recalls cover devices cleared under the same FDA submission (P030047); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Crack1
  • Detachment of Device or Device Component1

Patient problems

  • No Consequences Or Impact To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 24, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, THE 7MM X 30MM PRECISE PRO RAPID EXCHANGE (RX) CAROTID STENT SYSTEM COULDN'T BE DELIVERED TO THE TARGET SITE THEN THE DOCTOR FOUND THE TIP WAS RIPPED AND FRAGMENTED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS OPENED IN STERILE FIELD. THE DEVICE WAS CONTAMINATED BUT WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10896624 ↗
  • MalfunctionMar 30, 2017Crack; Crack

    DURING DEPLOYMENT OF THE PRECISE PRO RX US CAROTID SYSTEM STENT, THERE WAS SIGNIFICANT RESISTANCE AND THE STENT DELIVERY SYSTEM BROKE (THE DELIVERY SYSTEM CRACKED) WHILE ATTEMPTING TO APPROPRIATELY DELIVER THE STENT TO THE PROPER LOCATION. A DIFFERENT PRECISE PRO SYSTEM WAS USED TO COMPLETE THE PROCEDURE. DURING DEPLOYMENT THE PHYSICIAN FELT RESISTANCE AND THEN FELT NONE. THE STENT WAS NOT DEPLOYED BUT THE PART PINNED WAS AT THE ROTATING HEMOSTATIC VALVE (RHV). OUT AN ABUNDANCE OF CAUTION, THE PHYSICIAN REMOVED THE DEVICE AND DISCOVER THAT THE DELIVERY SYSTEM HAD BROKEN (CRACKED). THE PHYSICIAN TOOK A DIFFERENT PRECISE PRO CAROTID STENT AND COMPLETED THE CASE WITHOUT INCIDENT. THE PRODUCT WAS STORED, HANDLED, INSPECTED AND PREPPED ACCORDING TO THE INSTRUCTIONS FOR USE (IFU). THE DEVICE WAS PREPPED APPROPRIATELY. THE PHYSICIAN IS FAMILIAR WITH THE DEVICE AND PERFORMED ALL PREP TASK APPROPRIATELY. THERE WASN'T ANYTHING UNUSUAL NOTED ABOUT THE STENT DELIVERY SYSTEM PRIOR TO USE. THE INTENDED PROCEDURE/TARGET LESION WAS THE CAROTID ARTERY, HOWEVER THE LESION CHARACTERISTICS ARE UNKNOWN. WHEN REMOVED FROM THE TRAY, THE STENT WAS STILL CONSTRAINED WITHIN THE OUTER MEMBER/SHEATH. THE FOLL

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6446843 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P030047. One submission often covers many models, so check the product and lot details in each record.

  • Z-1257-2021 · Feb 11, 2021 · Completed
    PRECISE PRO RX CAROTID STENT implant system, Product Codes: PC0520RXC, PC0530RXC, PC0540RXC, PC0620RXC, PC0630RXC, PC0640RXC, PC0720RXC, PC0730RXC, PC0740RXC, PC0820RXC, PC0830RXC
  • Z-0729-2020 · Nov 14, 2019 · Open, Classified
    Cordis Precise Nitinol Stent System. The Cordis PRECISE Nitinol Stent System devices are designed to deliver a flexible, self-expanding endoluminal stent to the carotid vascula
  • Z-1947-2016 · May 4, 2016 · Terminated
    Cordis PRECISE (R) RX Nitinol Stent System (Biliary)
  • Z-1946-2016 · May 4, 2016 · Terminated
    Cordis PRECISE (R) PRO RX Nitinol Stent System (Carotid)

Other STENT, CAROTID devices

All STENT, CAROTID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page