Skip to content

S.M.A.R.T. RADIANZ SR08100XLRecalls & reported problems

Cordis US Corp. · S.M.A.R.T. RADIANZ Vascular Stent System 8x100mm 190CM · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 20705032082393), received between Aug 3, 2023 and Apr 29, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Defective Component” (1 report).

4 other recalls cover devices cleared under the same FDA submission (P020036, P120002); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Defective Component1
  • Detachment of Device or Device Component1
  • Difficult or Delayed Positioning1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AS REPORTED, AFTER IMPLANTATION OF AN 8MM X 100MM SMART RADIANZ SELF-EXPANDING STENT (SES), WHEN REMOVING THE DELIVERY SYSTEM FROM THE PATIENT OVER A 300CM VASSALLO FLOPPY GUIDEWIRE, IT WAS NOTICED THAT THE FRONT END WAS TORN OFF. THE 10CM SEPARATED SEGMENT REMAINED ON THE VASSALLO GUIDEWIRE AND WAS ABLE TO BE REMOVED WITHOUT ISSUE BY PULLING THE GUIDEWIRE. ADDITIONALLY, THERE WAS SOME RESISTANCE WHEN REMOVING THE DELIVERY SYSTEM DUE TO RESISTANCE WITH THE GUIDEWIRE. THE PROCEDURE CONTINUED BY IMPLANTING A 7MM X 60MM SMART VASCULAR SES AT THE EXTERNAL ILIAC ARTERY (EIA). THE RIGHT COMMON ILIAC ARTERY WAS THEN TREATED, AND THE PROCEDURE WAS COMPLETED. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT TWO SEPARATE LESIONS, ONE IN THE LEFT COMMON ILIAC ARTERY TO EXTERNAL ILIAC ARTERY (EIA), AND ONE IN THE RIGHT COMMON ILIAC ARTERY TO SUPERFICIAL FEMORAL ARTERY (SFA). THE EIA LESION HAD A STENOSIS OF 85% - 95% AND WAS MODERATELY CALCIFIED. THE SFA LESION HAD A STENOSIS OF 90%-100% AND WAS MODERATELY CALCIFIED. A 6F 90CM NON-CORDIS GUIDING SHEATH WAS USED FOR ACCESS FROM THE LEFT BRACHIAL ARTERY TO THE LEFT ILIAC ARTERY. VARIOUS UNKNOWN GUIDEWIRES WERE

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19201032 ↗
  • MalfunctionDec 7, 2023Difficult or Delayed Positioning; Difficult or Delayed Positioning

    AS REPORTED, WHEN USING AN 8MM X 100MM SMART RADIANZ SELF-EXPANDING STENT (SES) DELIVERY SYSTEM, THE STENT EDGE COULD BE DEPLOYED BUT THE STENT COULD NOT BE DEPLOYED. THE DEVICE WAS REPLACED WITH A NEW, UNKNOWN SMART CONTROL SES TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THIS WAS DURING A PROCEDURE TO TREAT AN ILIAC LESION. ADDITIONAL INFORMATION WAS REQUESTED BUT WAS NOT PROVIDED. THE DEVICE WILL NOT BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18286177 ↗
  • MalfunctionAug 3, 2023Defective Component; Defective Component

    AS REPORTED A SABERX RADIANZ PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY (PTA) DILATION CATHETER WAS USED BUT RUPTURED DUE TO SEVERE CALCIFICATION. ADDITIONALLY, A 6F SMART RADIANZ VASCULAR STENT WAS USED BUT THE STENT LENGTH WAS LONGER THAN EXPECTED. THE COMPLAINT DEVICES WERE REPLACED WITH TWO DIFFERENT SABERX RADIANZ AND SMART RADIANZ DEVICES OF DIFFERENT SIZES AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICES WERE BEING USED IN AN ENDOVASCULAR THERAPY (EVT) PROCEDURE. THE LENGTH OF THE LESION WAS ABOUT 8CM. THERE WAS SEVERE VESSEL TORTUOSITY WITH ONE-HUNDRED PERCENT (100%) STENOSIS. THERE WERE NO ISSUES EXPERIENCED WITH THE GUIDING SHEATH. THERE WERE NO VISIBLE SIGNS OF DEVICE/PACKAGE DAMAGE PRIOR TO USE. THERE WAS NO DIFFICULTY REMOVING THE PRODUCT FROM THE HOOP OR REMOVING THE PROTECTIVE BALLOON COVER NOR REMOVING THE STYLET OR ANY OF THE STERILE PACKAGING COMPONENTS. THERE WERE NO KINKS OR OTHER DAMAGES NOTED PRIOR TO INSERTING THE PRODUCT INTO THE PATIENT. THE DEVICE MAINTAINED NEGATIVE PRESSURE DURING PREPARATION. THERE WAS NO RESISTANCE/FRICTION WHILE INSERTING THE BALLOON INTO THE PATIENT. THERE WAS NO DIFFICULTY ADVANCING THE BALLOON CATHETER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17451859 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020036, P120002. One submission often covers many models, so check the product and lot details in each record.

  • Z-1510-2025 · Mar 4, 2025 · Open, Classified
    Cordis, REF C10040ML, Smart Control Vascular Stent System ILIAC, 10mm X 40mm 6F, SterileEO, Rx Only
  • Z-0728-2020 · Nov 14, 2019 · Open, Classified
    S.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy)
  • Z-1040-2012 · Jan 31, 2012 · Terminated
    S.M.A.R.T. CONTROL NITINOL STENT SYSTEM ILIAC: Self-Expanding Vascular Stents delivered by catheter for the indication of improving luminal diameter in patients with symptomatic a
  • Z-1556-2008 · Mar 10, 2008 · Terminated
    CORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atheroscler

Other STENT, SUPERFICIAL FEMORAL ARTERY devices

All STENT, SUPERFICIAL FEMORAL ARTERY devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page