S.M.A.R.T. RADIANZ SR07100XLRecalls & reported problems
Cordis Corporation · S.M.A.R.T. RADIANZ Vascular Stent System 7x100mm 190CM · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20705032082508), received between May 27, 2025 and Jan 9, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
4 other recalls cover devices cleared under the same FDA submission (P020036, P120002); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Difficult or Delayed Positioning1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2026Difficult or Delayed Positioning
AS REPORTED, THE STENT OF A 7MM X 100MM S.M.A.R.T. RADIANZ VASCULAR STENT SYSTEM DID NOT DEPLOY WHEN THE THUMB WHEEL WAS ROTATED. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS USED FOR A RIGHT EXTERNAL ILIAC ARTERY CHRONIC TOTAL OCCLUSION CASE VIA A RIGHT RADIAL ARTERY APPROACH. AFTER GUIDEWIRE PASSAGE, PRE-DILATION WAS PERFORMED, AND STENT DEPLOYMENT WAS ATTEMPTED; HOWEVER, THE THUMB WHEEL BECAME STIFF AND COULD NOT BE MOVED. DURING WITHDRAWAL OF THE DEVICE, STENT DEPLOYMENT BEGAN UNINTENTIONALLY, RESULTING IN DEPLOYMENT IN A PULLED-BACK STATE PROXIMAL TO THE TARGET LESION. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24021849 ↗MalfunctionMay 27, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, THE DELIVERY SYSTEM AND STENT OF A 7MM X 100MM 190CM S.M.A.R.T. RADIANZ VASCULAR STENT SYSTEM BROKE APART WHILE THE OPERATOR USED IT ON THE PATIENT. THERE WERE NO REPORTS OF PATIENT INJURY. THE DEVICE WAS STORED AS PER LABELING. THE DEVICE WAS OPENED IN A STERILE FIELD. ADDITIONAL INFORMATION WAS REQUESTED BUT IS NOT AVAILABLE. THE DEVICE WILL BE RETURNED FOR EVALUATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22088679 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020036, P120002. One submission often covers many models, so check the product and lot details in each record.
- Z-1510-2025 · Mar 4, 2025 · Open, ClassifiedCordis, REF C10040ML, Smart Control Vascular Stent System ILIAC, 10mm X 40mm 6F, SterileEO, Rx Only
- Z-0728-2020 · Nov 14, 2019 · Open, ClassifiedS.M.A.R.T. Nitinol Stent System and the S.M.A.R.T. Control Nitinol Stent System. The SMART stent system includes a self-expanding stent made of Nitinol (nickel-titanium alloy)
- Z-1040-2012 · Jan 31, 2012 · TerminatedS.M.A.R.T. CONTROL NITINOL STENT SYSTEM ILIAC: Self-Expanding Vascular Stents delivered by catheter for the indication of improving luminal diameter in patients with symptomatic a
- Z-1556-2008 · Mar 10, 2008 · TerminatedCORDIS S.M.A.R.T. Control" Nitinol Stent System Model Number(s): C10040SL, Manufactured in Mexico. Indicated for improving luminal diameter in patients with symptomatic atheroscler
Other STENT, ILIAC devices
- Cordis S.M.A.R.T. Control Vascular Stent System C10040MLCordis US Corp. · STENT, ILIAC1 recall3 reports
- Absolute Pro 1012534-100Abbott Vascular Inc. · STENT, ILIAC51 reports
- Protégé™ GPS™ SERB65-09-60-80Ev3, Inc · STENT, ILIAC1 recall0 reports
- Cordis S.M.A.R.T. Control Vascular Stent System C09060MLCordis US Corp. · STENT, ILIAC1 recall0 reports
- Absolute Pro 1012535-100Abbott Vascular Inc. · STENT, ILIAC1 death41 reports
- Omnilink Elite 1012631-39Abbott Vascular Inc. · STENT, ILIAC27 reports
- Omnilink Elite 1012632-29Abbott Vascular Inc. · STENT, ILIAC25 reports
- Absolute Pro 1012536-100Abbott Vascular Inc. · STENT, ILIAC21 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page