NIM TriVantage® 8229736 (UDI 20763000041513)Recalls & reported problems
Medtronic Xomed, Inc. · EMG TUBE 8229736 3PK NIM TRIVANTAGE 6MM · Class 2
All models of NIM TriVantage (18)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20763000041513), received between Jul 16, 2025 and Jul 21, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).
2 FDA recall records name this device's identifier, 2 still open.
1 other recall cover devices cleared under the same FDA submission (K112686); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 7, 2023Open, Classified
Firm received reports of customers experiencing noise from the NIM System, lead-off or high impedance issues, or loss of/intermittent nerve monitoring,
Z-0696-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
- Dec 22, 2021Open, Classified
There is potential for multiple issues due to a manufacturing nonconformity.
Z-0743-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings2
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Insufficient Information1
- Unintended Extubation1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 21, 2026Incorrect, Inadequate or Imprecise Result or Readings
Additional information received, there was a procedural delay of 10 minutes for troubleshooting. There was no patient injury.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions; Unintended Extubation
MAUDE report 25884749 ↗MalfunctionJul 16, 2025Signal Artifact/Noise; Incorrect, Inadequate or Imprecise Result or Readings; Signal Artifact/Noise; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING THYROIDECTOMY PROCEDURE, THERE WAS EXCESS NOISE HAPPENING THROUGHOUT NIM VITAL CASE USING SIZE 6 TRIVANTAGE TUBE. NIM VITAL SOFTWARE HAS ALREADY BEEN UPDATED. PASSED ELECTRODE CHECK AT BEGINNING OF CASE AND THEN EXCESS NOISE STARTED TO HAPPEN WITHIN THE FIRST 30 MIN OF CASE. WENT THROUGH ALL TROUBLESHOOTING. PATIENT WAS NOT LIGHT, BITE BLOCK WAS PLACED, MACHINE PLUGGED INTO SEPARATE OUTLET, INCREASED THRESHOLD TO 150, RESTARTED MACHINE, SWITCHED OUT PI BOX. WHEN PERFORMING ELECTRODE CHECK VOCALIS 2 IS TURNING RED THEN GREEN THEN BACK TO RED (CONSTANTLY CIRCLING AND NEVER STAYING GREEN). PHYSICIAN WAS EXTREMELY UPSET AS THIS WAS A HUGE DISTRACTION DURING THE CASE. STAFF WAS ALSO UPSET AND FRUSTRATED WITH EXCESS NOISE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22517340 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112686. One submission often covers many models, so check the product and lot details in each record.
- Z-1420-2013 · Mar 4, 2013 · TerminatedREF 8229705 x1 NIM TriVantage EMG Endotracheal Tube 5.0 mm I.D. x 6.5 mm O.D. Medtronic Xomed, Inc., 6743 Southpoint Drive North, Jacksonville, FL 32216-0980 USA STERILE Rx Only
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page