Skip to content

NIM TriVantage® 8229736 (UDI 20763000041513)Recalls & reported problems

Medtronic Xomed, Inc. · EMG TUBE 8229736 3PK NIM TRIVANTAGE 6MM · Class 2

All models of NIM TriVantage (18)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 20763000041513), received between Jul 16, 2025 and Jul 21, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K112686); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 7, 2023Open, Classified

    Firm received reports of customers experiencing noise from the NIM System, lead-off or high impedance issues, or loss of/intermittent nerve monitoring,

    Z-0696-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 22, 2021Open, Classified

    There is potential for multiple issues due to a manufacturing nonconformity.

    Z-0743-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings2
  • Signal Artifact/Noise1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Insufficient Information1
  • Unintended Extubation1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2026Incorrect, Inadequate or Imprecise Result or Readings

    Additional information received, there was a procedural delay of 10 minutes for troubleshooting. There was no patient injury.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions; Unintended Extubation

    MAUDE report 25884749 ↗
  • MalfunctionJul 16, 2025Signal Artifact/Noise; Incorrect, Inadequate or Imprecise Result or Readings; Signal Artifact/Noise; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING THYROIDECTOMY PROCEDURE, THERE WAS EXCESS NOISE HAPPENING THROUGHOUT NIM VITAL CASE USING SIZE 6 TRIVANTAGE TUBE. NIM VITAL SOFTWARE HAS ALREADY BEEN UPDATED. PASSED ELECTRODE CHECK AT BEGINNING OF CASE AND THEN EXCESS NOISE STARTED TO HAPPEN WITHIN THE FIRST 30 MIN OF CASE. WENT THROUGH ALL TROUBLESHOOTING. PATIENT WAS NOT LIGHT, BITE BLOCK WAS PLACED, MACHINE PLUGGED INTO SEPARATE OUTLET, INCREASED THRESHOLD TO 150, RESTARTED MACHINE, SWITCHED OUT PI BOX. WHEN PERFORMING ELECTRODE CHECK VOCALIS 2 IS TURNING RED THEN GREEN THEN BACK TO RED (CONSTANTLY CIRCLING AND NEVER STAYING GREEN). PHYSICIAN WAS EXTREMELY UPSET AS THIS WAS A HUGE DISTRACTION DURING THE CASE. STAFF WAS ALSO UPSET AND FRUSTRATED WITH EXCESS NOISE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22517340 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112686. One submission often covers many models, so check the product and lot details in each record.

  • Z-1420-2013 · Mar 4, 2013 · Terminated
    REF 8229705 x1 NIM TriVantage EMG Endotracheal Tube 5.0 mm I.D. x 6.5 mm O.D. Medtronic Xomed, Inc., 6743 Southpoint Drive North, Jacksonville, FL 32216-0980 USA STERILE Rx Only

Other STIMULATOR, NERVE devices

All STIMULATOR, NERVE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page