NIM 8225825XRecalls & reported problems
Medtronic Xomed, Inc. · PROBE 8225825X 3PK INCREMT STD PRASS TIP · Class 2
All models of NIM (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20763000112459), received between May 1, 2024 and Apr 13, 2026 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “False Negative Result” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- False Negative Result3
- Inappropriate Audible Prompt/Feedback2
- Incorrect, Inadequate or Imprecise Result or Readings2
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Hemorrhage/Blood Loss/Bleeding1
- Nerve Damage1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 13, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT INTRA OPERATIVELY, THAT THE INCREMENTING PROBE IS NOT FUNCTIONING. THE PROBE WAS STILL STIMULATING AND THEY WERE STILL RECEIVING POSITIVE NERVE CONFIRMATION HOWEVER, THE INCREMENTING PORTION WAS NOT FUNCTIONING. THE PROBE GAVE READINGS OF 1.0. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24859783 ↗MalfunctionNov 4, 2025False Negative Result; False Negative Result
ADDITONAL INFORMATION RECEIVED THAT NO ELECTROSURGICAL EQUIPMENT WAS USED DURING THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23460935 ↗MalfunctionJan 6, 2025Inappropriate Audible Prompt/Feedback; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT INTRA OP, EXCESSIVE ARTIFACTS BEING PICKED UP BY THE NIM, SO THE SURGEON REQUESTED FOR A SECOND ONE TO BE OPENED, WHICH WORKED FINE. NO CONSEQUENCES OR IMPACT TO PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21077404 ↗InjuryMay 6, 2024False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings; False Negative Result; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED THAT DURING THE TOTAL THYROID FOR GRAVES DISEASE AND GOITER, STARTED THE PROCEDURE ON LEFT SIDE LOBECTOMY, NIMS TUBE FOR INTUBATION AND MONITORING FOR PROCEDURE. UNIT SETUP AND TESTED AT PROCEDURE BEGINNING, AND RECEIVING FEEDBACK DURING BEGINNING OF PROCEDURE. LATER IN THE PROCEDURE, THE SURGEON STATED "NIMS WAS NOT WORKING". AT THAT TIME, IT WAS ON THE MONITORING SCREEN, WITH NO ERRORS OR MESSAGES, AND PROVIDING NO FEEDBACK. RESTARTED THE UNIT AND HARDWIRED THE PATIENT INTERFACE WITH THE UNIT. AT THIS POINT IT HAD INTERMITTENT "LOST CONNECTION" MESSAGES. CONNECTIONS WERE DOUBLE CHECKED AND THEN MOVED TO RIGHT SIDE AND ALL WORKED FINE. AT PROCEDURE END, SURGEON STATED COMPLICATION OF PROCEDURE WAS TRANSECTED NERVE. THERE WAS A BLOOD LOSS ABOUT 500ML. THE PATIENT COULD SWALLOW BUT HIS VOICE IS BREATHY; UPON FOLLOW-UP THE PATIENT WAS HOARSE, AND VOICE HAD NOT RECOVERED SEVERAL DAYS LATER.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Nerve Damage
MAUDE report 19249239 ↗MalfunctionMay 1, 2024False Negative Result; Inappropriate Audible Prompt/Feedback; False Negative Result; Inappropriate Audible Prompt/Feedback
IT WAS REPORTED THAT DURING PAROTID CASE, THE SURGEON HAD A NEGATIVE TONE (TWEEDLE) WHEN TOUCHING A NERVE. PHYSICIAN EXPECTED TO HEAR THE POSITIVE ¿BUMP BUMP BUMP¿ BIPHASIC WAVE SOUND INDICATING THE NERVE WAS BEING STIMULATED. INSTEAD, THE TWEEDLE INDICATING THAT THEY WERE NOT TOUCHING NERVE RESPONDED. PHYSICIAN WAS CONFIDENT HE WAS ON NERVE. OLDER PHYSICIAN THAT WAS TRAINED WITHOUT THE NIM AND VERY STRONG ANATOMY KNOWLEDGE. FELT STRONGLY THAT THE SYSTEM WAS INCORRECT. THE PROCEDURE WAS EXTENDED FOR LESS THAN 1 HOUR. THERE WAS NO PATIENT IMPACT. EVENT COULD NOT BE REPLICATED BY REP OR BIOMED. USED SUBSEQUENTLY WITH DIFFERENT PI BOX AND WORKED AS EXPECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19219366 ↗
Other STIMULATOR, NERVE devices
- NIM Vital NIM4CM01Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls260 reports
- NIM TriVantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls176 reports
- NIM® 8229306Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall1 death167 reports
- NIM Vital™ NIM4CPB1Medtronic Xomed, Inc. · STIMULATOR, NERVE3 recalls45 reports
- NIM Trivantage® 8229707Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall142 reports
- NIM® 8229307Medtronic Xomed, Inc. · STIMULATOR, NERVE1 recall138 reports
- NIM TriVantage® 8229706Medtronic Xomed, Inc. · STIMULATOR, NERVE2 recalls78 reports
- NIM-Response® 8253002Medtronic Xomed, Inc. · STIMULATOR, NERVE171 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page