Skip to content

CD Horizon™ ModuLeX™ Spinal System 559200028 (UDI 20763000608211)Recalls & reported problems

Medtronic Sofamor Danek, Inc. · MAS/SS PK 559200028 STERILE 5.5/6.0 2PK · Class 2 · Implant

All models of CD Horizon ModuLeX Spinal System (34)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 20763000608211), received between Jun 25, 2025 and Aug 26, 2025 — classified as 5 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 4 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated5

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING IDIPATHIC SCOLIOSIS CASE-OPEN MANIPULATION FOR IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT IT WAS NOTICED IN THE 1 YEAR FOLLOW UP X RAY, THAT THE SET SCREWS HAVE COME OFF THE DISTAL HEADS OF 3-4 DISTAL SCREWS AT L2/L3 LEVELS. THE REVISION SURGERY IS SCHEDULED FOR (B)(6) 2025. IT WAS LATER CONFIRMED THAT 5 SET SCREWS WERE EXPLANTED IN THE REVISION SURGERY AND ARE AVAILABLE FOR RETURN. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892343 ↗
  • InjuryAug 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING IDIPATHIC SCOLIOSIS CASE-OPEN MANIPULATION FOR IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT IT WAS NOTICED IN THE 1 YEAR FOLLOW UP X RAY, THAT THE SET SCREWS HAVE COME OFF THE DISTAL HEADS OF 3-4 DISTAL SCREWS AT L2/L3 LEVELS. THE REVISION SURGERY IS SCHEDULED FOR (B)(6) 2025. IT WAS LATER CONFIRMED THAT 5 SET SCREWS WERE EXPLANTED IN THE REVISION SURGERY AND ARE AVAILABLE FOR RETURN. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892295 ↗
  • InjuryAug 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING IDIPATHIC SCOLIOSIS CASE-OPEN MANIPULATION FOR IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT IT WAS NOTICED IN THE 1 YEAR FOLLOW UP X RAY, THAT THE SET SCREWS HAVE COME OFF THE DISTAL HEADS OF 3-4 DISTAL SCREWS AT L2/L3 LEVELS. THE REVISION SURGERY IS SCHEDULED FOR (B)(6) 2025. IT WAS LATER CONFIRMED THAT 5 SET SCREWS WERE EXPLANTED IN THE REVISION SURGERY AND ARE AVAILABLE FOR RETURN. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892285 ↗
  • InjuryAug 26, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING IDIPATHIC SCOLIOSIS CASE-OPEN MANIPULATION FOR IDIOPATHIC SCOLIOSIS. IT WAS REPORTED THAT IT WAS NOTICED IN THE 1 YEAR FOLLOW UP X RAY, THAT THE SET SCREWS HAVE COME OFF THE DISTAL HEADS OF 3-4 DISTAL SCREWS AT L2/L3 LEVELS. THE REVISION SURGERY IS SCHEDULED FOR (B)(6) 2025. IT WAS LATER CONFIRMED THAT 5 SET SCREWS WERE EXPLANTED IN THE REVISION SURGERY AND ARE AVAILABLE FOR RETURN. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22892233 ↗
  • InjuryJun 25, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE STATING THAT THE REVISION SURGERY IS SCHEDULED FOR (B)(6) 2025, AND HE CAN'T PROVIDE ANY FURTHER INFORMATION UNTIL THEN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22325193 ↗

Other Appliance, fixation, spinal intervertebral body devices

All Appliance, fixation, spinal intervertebral body devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page